Page 1
Instructions for Use Operating Manual Rigid Laryngostroboscope, Models 9106 and 9108 PENTAX of America, Inc. 3 Paragon Drive Montvale, NJ 07645-5297 USA Toll-Free Telephone: 1-800-431-5880 (In USA and Canada only Tel: (201) 571-2300 Service: (201) 391-4189 email: customerservice@pentaxmedical.com Web: www.pentaxmedical.com/us...
Page 2
While technical inaccuracies, typographical errors, and editorial omissions may occur, PENTAX of America, Inc. disclaims all direct, indirect or consequential damages that may result from such errors. PENTAX of America, Inc. will make every effort to correct the deficiencies in new editions and updates.
Table of Contents Introduction ................4 1-1 Intended Use................4 1-2 Warnings and Precautions ............4 1-3 Using this Manual ..............5 Specifications ................. 5 2-1 Components ................5 2-2 Appearance ................6 2-3 Specifications ................6 2-4 Environmental and Storage Conditions ........7 2-5 Prolonging the Life of the Rigid Laryngostroboscope ....
1-1 Intended Use The Rigid Laryngostroboscopes, Models 9106 and 9108, are intended to observe glottic action with the use of a stroboscopic light source. The instrument is inserted perorally when indications consistent with the need for observation are observed.
Italics Reference to other sections in this manual or to other documents. Specifications 2-1 Components The 9106 and 9108 include the following components: • Rigid Laryngostroboscope • PENTAX Medical Laryngeal Stroboscope Adaptor, Model 7175-4200 D04-IFU-0001400 Revision: B...
2-3 Specifications Total Length 252 mm Working Length 189 mm Outer Diameter 9.5 mm Actual Visual Field Angle 35 degree 9106 Specifications Angled View 70 degree Depth of Field 2 mm-40 mm Integral Light Cable Light Guide Length 1727 mm...
Atmosphere pressure: 700mb to 1060mb (0.69atm to 1.05atm) 2-5 Prolonging the Life of the Rigid Laryngostroboscope The 9106 and 9108 provide high-quality performance over many examinations. The following precautions are essential to guarantee their longevity: • Bend the fiberoptic cable as little as possible. Never kink the cable.
You must loosen the screw prior to removing the adaptor from the eyecup in order to avoid damage to the rigid Laryngostroboscope. Note To use the 9106 or 9108 with other light sources, adaptors may be required. Please contact your local representative for information.
3-2 Cleaning, High Level Disinfection, and Sterilization Warning The 9106 and 9108 are provided in a non-sterilized condition. Do not use until you clean, disinfect and sterilize the rigid Laryngostroboscope, as this could result in bodily harm to the patient.
After completing the soaking procedure, remove the devices from the basin of detergent solution and rinse them thoroughly under lukewarm, running tap water for a minimum of one minute. Clean the devices with a general purpose detergent for healthcare facilities, prepared in accordance with the manufacturer’s recommendations.
• This device is compatible with STERRAD 100S when following the ® sterilant manufacturer’s procedures and guidelines. 3-3 Use Caution Thoroughly inspect the device before each use. Check the function of the device and visually inspect optics for moisture or contamination.
Electronic Equipment, this symbol indicates that the product must not be disposed of as unsorted waste, but should be collected separately. Contact your local PENTAX distributor for correct disposal and recycling. By disposing of this product correctly you will help ensure that the...