Purging of water from lumens
10) Using a syringe filled with air, flush the lumen of the suction valve (OF-B179) to
purge residual water. (see Figure 1.33, p41)
1-3-2-4. Drying
WARNING
Regardless of the quality of the rinse water used, it is essential to perform a
final alcohol rinse followed by forced air in order to completely dry the lumens
of components and accessories, and prevent bacterial colonization and/or
infections associated with waterborne microorganisms.
Flushing of lumens with alcohol
1)
Using a syringe filled with 70-90% medical grade ethyl or isopropyl alcohol, flush
the lumen of the suction valve (OF-B179) with 2 mL of alcohol. (see Figure 1.33,
p41)
Purging of alcohol from lumens
2)
Using a syringe filled with air, flush the lumen of the suction valve (OF-B179) with
air to purge residual alcohol. (see Figure 1.33, p41)
Drying of all external surfaces
3)
Gently dry all external surfaces of components and accessories with a sterile gauze
moistened with 70-90% medical grade ethyl or isopropyl alcohol.