1-1-5. Inspection of Reprocessing Accessories

Before use, inspect reprocessing accessories according to the following procedure.
WARNING
• Replace reprocessing accessory with a new one when inspection of the
device indicates that it is damaged or unable to function properly.
1-1-5-1. Inspection of PVE Soaking Cap (OE-C9)
1)
Check that there is no cracking on the outer surface of the PVE Soaking Cap.
2)
Check that there are no scratches, cracking, or chipping of the sealing surfaces inside
the PVE Soaking Cap .
Figure 1.3
1-1-5-2. Inspection of Ventilation Cap (OF-C5)
1)
Make sure that of the Locking Groove Potion of the Ventilation Cap is not deformed.
2)
Check that there are no scratches, cracking, or chipping of the O-ring inside the
Ventilation Cap.
Figure 1.4