Pentax EB-1970TK Instructions For Use Manual page 53

Video bronchoscopes
Table of Contents

Advertisement

1-3-3-1-2. Wrapping
1)
Prior to steam sterilization, wrap the components and accessories individually with
two layers of one-ply Kimguard KC200 sterilization wrap (Kimberly-Clark) using
sequential wrapping technique.
1-3-3-1-3. Parameters
Steam sterilization can be performed under the following conditions:
Sterilizer Type
Pre-vacuum
NOTE
• Validation testing of these parameters was performed using two layers of
one ply Kimberly Clark Wrap KC-200 Sterilization Wrap (Kimberly-Clark).
Sequential wrapping technique was employed.
• Biological and/or chemical indicators used must be appropriate for the
stated sterilization cycle parameters and cleared by FDA.
1-3-3-2. Sterilization using STERRAD
The suction control valve (OF-B179) and the inlet seal (OF-B190) can be sterilized using
®
the STERRAD
NX™ sterilization.
1-3-3-2-1. Items required
• protective garments such as gloves, gowns, face masks, etc., to minimize the
risk of cross contamination.
• APTIMAX Instrument Tray (270 mm in width, 576 mm in length, 100 mm in height)
• APTIMAX Instrument Tray Mat (254 mm in width, 546 mm in length)
• instrument wrap recommended by STERRAD
system
• tape recommended by STERRAD
Temperature
Exposure
132°C
4 minutes
135°C
3 minutes
®
NX™ system
®
for use of STERRAD
– 48 –
Drying Time
30 minutes
16 minutes
®
®
for use of STERRAD
®
NX™ system
1
NX™
Accessory
Sterilization