Immunity test
Conducted RF IEC/EN
61000-4-6
Radiated RF IEC/EN
61000-4-3
is the AC mains voltage prior to application of the test level.
U
T
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
IEC/EN 60601
Test levels
3 Vrms
3 Vrms
150 kHz to 80 MHz
3 V/m
3 V/m
80 MHz to 2.5 GHz
Compliance level
Portable and mobile RF communications
equipment should be used no closer to
any part of the MD series LED Display,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2
d = 1.2
d = 2.3
Where "P" is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and "d" is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with symbol:
Electromagnetic environment –
guidance
P
P
80 MHz to 800 MHz
P
800 MHz to 2.5 Ghz
Regulatory statements
57