4-7.

Testing Amps Performance

Perform the following procedure to test the amps funtion of the Tester.
Ensure that the calibrator is in standby mode before making
any connection between the calibrator and Tester. dangerous
voltages may be preSent on the leads and connectors.
1. Press E and turn on the Tester to enter the Setup screen.
2. Using E and < > , select "OTHER" for Clamp Setting.
3. Press E to exit the Setup screen.
4. Connect a cable from the Output Voltage HI and LO connectors of calibrator "A" to
the Current Probe connector on the Tester.
5. Press M until the text screen is displayed on the Tester.
6. Press F until A is displayed in the upper right corner of the display.
Table 4-2. Volts Performance
V rms
Range
Min
20
50
100
200
500
500
1000
1000
20
4.5
5.5
50
14.7
15.3
100
64.4
65.6
200
129.1
130.9
500
205
211
500
346
354
1000
475
485
1000
595
605
Warning
Performance Testing and Calibration
Performance Limits
V pk
Max
Min
Max
6.5
7.7
20.5
22.0
89.8
94.1
179.9
187.8
285
303
483
507
662
695
828
868
Performance Tests
V dc
Min
Max
4.4
5.6
14.4
15.6
63.4
66.6
127.1
132.9
201
215
340
360
467
493
585
615
4
4-5