Introduction; Required Equipment - Fluke 39 Service Manual

Power meter & power harmonics analyzer
Hide thumbs Also See for 39:
Table of Contents

Advertisement

Service procedures in this chapter should be performed by
qualified service personnel only. To avoid electrical shock, do
not perform any servicing unless you are qualified to do so.

4-1. Introduction

This chapter provides calibration and performance tests that allow you to verify that the
Tester is operating within published specifications.

4-2. Required Equipment

The equipment in Table 4-1 is required for performance testing and calibration. If the
recommended models are not available, equipment with equivalent specifications should
be used.
Equipment Type
Calibrator A, volts
DC Voltage:
AC Voltage:
Calibrator B, volts*
AC Voltage:
Phase meter*
Phase accuracy: 0.5 degrees @ 60 Hz
comparing 10 mV and 5 V inputs
Dual channel
AC Voltage:
signal generator*
* A second calibrator, phase meter and dual channel signal generator is not required when using one
Fluke 5500A Multi-Product Calibrator.
Warning
Table 4-1. Recommended Test Equipment
Minimum Specifications
Range: 1 mV - 600V
Accuracy: 0.12%
Range:1 mV - 600V
Frequency: 60 Hz
Accuracy: 0.12%
Range: 30 mV-1.4V
Frequency: 60 Hz
Accuracy: 0.12%
Range: 50 mV - 20 V p-p
Frequency: 60 Hz
Accuracy: 0.5%
Phase accuracy: 0.5 degrees @ 60 Hz
sourcing 10 mV and 5 V inputs
Performance Testing and Calibration
Recommended Model
Fluke 5700A
Fluke 5500A
Fluke 5200A
Fluke 5500A
Clarke-Hess Model 6000A
Fluke 5500A
Fluke 5500A
4

Introduction

4-3