Operating Instructions; Specifications - Fluke 39 Service Manual

Power meter & power harmonics analyzer
Hide thumbs Also See for 39:
Table of Contents

Advertisement

The following additional accessories are supplied with the Fluke 41B:
RS-232 Cable
9-Pin to 25-Pin Adapter
Plug Adapter
3.5 inch Micro-Floppy Disk (FlukeView™ 41 Software)
Optional accessories for both the Fluke 39 and 41B are as follows:
80i-1000s AC Current Probe
C41s
Soft Carrying Case
1-8.

Operating Instructions

Operating instructions for the Fluke 39 and 41B can be found in the Users Manual
(Fluke PN 942847). See How to Obtain Parts on page 6-1.

1-9. Specifications

Accuracy is specified for a period of one year after calibration.
Specifications expressed in the "±(2% + 3 digits)" format are saying that the error is a percent of
reading plus a number of least significant digits. The least significant digit is the smallest value that
can be displayed in a range and can be found in the resolution section of the Range and
Resolution specifications table.
When a harmonic error is expressed in the "13
is indicating that there is a linear increase in error from the lower harmonic to the upper harmonic.
The error can be determined as a ratio of the desired harmonic to the number of harmonics in the
expression times the difference in error over the range plus the least error. For example, to find the
nd
22
Harmonic error from the above equation:
((Desired Harmonic - Lower Harmonic) / (Upper Harmonic - Lower Harmonic)) X Error Difference +
Least Error
((22 - 13) / (31 - 13)) X 3 + 2 = 3.5%
Therefore, the error for the 22
Frequency Range, Fundamental:
Minimum Input Levels:
Volts Measurements (True rms):
Input Range:
Basic Accuracy*:
rms (ac + dc):
peak, dc:
* < 15V rms, add 2 digits
Input Impedance:
Crest Factor:
Amps Measurements (True rms)
(1 mV/A) Isolated Input
th
(±(2% + 2 digits)) --- 31
harmonic is ±(3.5% + 2 digits).
nd
6-65 Hz and dc
5V rms or 1A rms
5.0V to 600V rms (ac + dc)
5.0V to ±933V peak
±(0.5% + 2 digits); 6-65 Hz
±(2% + 3 digits)
1 MΩ, balanced
> 3.0 below 300V, 1.56 @ 600V
Introduction and Specifications
Specifications
st
(±(5% + 2 digits))" form, it
1
1-5