Troubleshooting - Konica Minolta JM-103 User Manual

Jaundice meter
Hide thumbs Also See for JM-103:
Table of Contents

Advertisement

Troubleshooting

WARNING:
Only facility-authorized personnel should troubleshoot the
Jaundice Meter. Troubleshooting by unauthorized personnel
could result in personal injury or equipment damage.
If an abnormality occurs with the Jaundice Meter, perform the
following:
1.
Refer to the table below, and take the necessary action given.
2.
If the abnormality still appears, set the Power switch to the Off
position, and then return it to the On position.
3.
If the abnormality still continues, contact the nearest authorized
service facility.
Symptom
The display is blank
when the Power switch
is in the On position.
The display suddenly
goes blank during a
measurement.
The Charger lamp
does not illuminate
when the Jaundice
Meter is placed on the
charger unit.
It is impossible to take
measurements.
Page 7 - 2
Troubleshooting
Possible Cause
The batteries are
exhausted.
The batteries are
exhausted.
The charger unit and
the AC adapter are not
plugged into an AC
outlet correctly.
The Jaundice Meter is
not placed in the
charger unit correctly.
The Ready lamp is not
illuminated.
The batteries are
exhausted.
Jaundice Meter (Model JM-103) User Manual (usr070rb)
Action
Charge the battery (see
"Charging the Battery"
on page 4-1).
Charge the battery (see
"Charging the Battery"
on page 4-1).
Plug the charger unit
and the AC adapter into
an appropriate power
source correctly.
Place the Jaundice
Meter perpendicular to
the measuring point in
the charger unit.
Before taking a
measurement, ensure
that the Ready lamp is
illuminated.
Charge the battery (see
"Charging the Battery"
on page 4-1).