Konica Minolta JM-103 User Manual page 29

Jaundice meter
Hide thumbs Also See for JM-103:
Table of Contents

Advertisement

5.
Place the measuring probe
perpendicular to the checker, and push
it gently until a click sounds.
6.
If the measuring probe contacts the
checker at an angle, place it
perpendicular, and take the
measurement again.
NOTE:
The display interchanges between the L-
value, the measured value of the long-
optical path, and the S-value, the measured
value of the short-optical path. When the L-
value is displayed, "•" appears in the upper
left-hand corner of the display.
7.
Confirm the measured value. If both
the L-value and the S-value are within
± 1.0 of the reference values indicated
on the checker, the values are
acceptable.
8.
If the measured value exceeds ± 1.0 of
the reference value, perform the
following:
a.
Clean both the checker and the
measuring probe.
b.
Place the measuring probe perpendicular to the checker, and
push it gently until a click sounds.
c.
If the measured value still exceeds ± 1.0 of the reference value,
contact the nearest authorized service facility.
9.
Close the cover of the checker.
10. Set the Power switch to the Off position.
Jaundice Meter (Model JM-103) User Manual (usr070rb)
Page 4 - 5