Process documentation

48
Each washer-disinfector is supplied with a 5 m long interface cable
for data transfer between the Profitronic system and an external re-
port printer or PC.
The interface cable is rolled up inside the machine and must only
be connected by Miele Service.
The serial interface is RS 232 compatible. For the interface configura-
tions see "PC/Printer Function" in the Programming manual.
Various printers can be used as external printers:
– Miele PRT 100 report printer
– Epson-compatible (contact Miele for a list of suitable models).
Pin configuration in the 9-pole sub-D connector on the interface
cable:
5 GND (base)
3 TXD (transmit)
2 RXD (receive)
1-4-6 (linked)
7-8 (linked)
A standard null-modem or laplink cable can be connected.
The extension cable to the printer/PC must not exceed 10 m in
length.
Please note the following when connecting a printer or PC:
– Only use an industry-standard PC or printer (e.g. in Germany TÜV,
or VDE approved).
– The size of the printer or PC must be taken into account when in-
stalling the machine.
The report printer settings are described in detail in the Programming
manual under "PC/Printer function".
The following reports are available:
1. Programme output
2. Additional function output
3. Wash report output
4. Error log output