Kodak i5000 Series User Manual page 44

Hide thumbs Also See for i5000 Series:
Table of Contents

Advertisement

Performing a Patch test
Performing a UDDS
calibration
A-61670 August 2014
Use the Patch test to verify that the patch sheets can be detected.
1. Touch Patch Test. The Patch Test screen will be displayed.
2. Place a document(s) with a patch(es) in the input elevator.
3. Press the Start/Resume button. When the document(s) has been
scanned, the results of the patch test will be displayed on the touchscreen.
The results show the number of patches and patch types that were
recognized by the scanner
4. Touch Cancel to return to the Idle screen.
UDDS calibration is used to calibrate the ultrasonic document detection
system for multifeeds. UDDS calibration is rarely needed. This calibration
should only be performed if there is a feeding or multifeed detection issue.
1. Touch Calibrate. The Calibration screen will be displayed.
2. Touch UDDS Calibration to display the UDDS Calibration screen.
3. You will be prompted to place a 20 lb. or 80 g/m
the input elevator in landscape orientation.
4. Press the Start/Resume button. The calibration results will be displayed.
5. When finished, touch Cancel to return to the Diagnostic screen.
6. Touch Close to return to the Idle screen.
.
2
bond sheet of paper in
3-21