Rtsi Trigger Lines; Figure 4-25. Daqarb 5411 Rtsi Trigger Lines And Routing - National Instruments DAQArb 5411 User Manual

High-speed arbitrary waveform generator
Hide thumbs Also See for DAQArb 5411:
Table of Contents

Advertisement

RTSI Trigger Lines

© National Instruments Corporation
The sample clock for integral subdivisions of 40 MHz will always have
a high pulse width of 25 ns. If the t
time of your device, then you can use the falling edge of the sample
clock output (PCLK) to register the digital pattern data.
The DAQArb 5411 contains seven trigger lines and one RTSI clock line
available over the RTSI bus to send and receive DAQArb 5411-specific
information to other boards having RTSI connectors. Figure 4-25 shows
the RTSI trigger lines and routing of DAQArb 5411 signals to the RTSI
switch.
SYNC
Start Trigger
Marker
RTSI Trigger
Board Clock
Master/Slave
RTSI Clock

Figure 4-25. DAQArb 5411 RTSI Trigger Lines and Routing

For phase locking to other boards as a master, the 5411 sends an
onboard 20 MHz signal to the RTSI Osc line as a Board Clock signal.
For locking to other devices as a slave, the DAQArb 5411 receives the
RTSI Osc line as a RTSI Clock signal.
The DAQArb 5411 can receive a hardware trigger from another board
as a RTSI trigger signal on any of the RTSI trigger lines, RTSI <0.. 6>.
The marker generated during waveform generation in Arb mode can be
routed to any of the RTSI bus trigger lines.
The trigger generated on the DAQArb 5411 can be routed to other
boards through any of the RTSI bus trigger lines.
time is insufficient for the hold
co
RTSI 0
RTSI 1
RTSI 2
RTSI 3
RTSI
RTSI 4
Switch
RTSI 5
RTSI 6
RTSI Osc