Specifications - Fluke 175 User Manual

Table of Contents

Advertisement

175, 177, 179
Users Manual
Table 8. Replacement Battery and Fuses
F1
Item
W F1 Fuse, 440 mA, 1000 V, FAST
W F2 Fuse, 11 A, 1000 V, FAST
B1 Battery, 9 V Alkaline
NEDA 1604 / 1604A
18
B1
F2
aik13f2.eps
Part Number
943121
803293
614487

Specifications

Accuracy is specified for 1 yr after calibration, at operating
temperatures of 18 °C to 28 °C, with relative humidity at 0 % to
90 %. Accuracy specifications take the form of: ±([% of Reading] +
[Counts])
Maximum voltage between any
terminal and earth ground ................. 1000 V
W Fuse Protection for mA inputs ..... 0.44 A, 1000 V, IR 10 kA
W Fuse Protection for A input .......... 11 A, 1000 V, IR 17 kA
Display ................................................ Digital: 6000 counts, updates 4/sec
Bar Graph........................................ 33 segments, Updates 40x/sec
Frequency ....................................... 10,000 counts
Capacitance .................................... 1000 counts
Altitude
Operating ........................................ 2000 m
Storage ........................................... 12,000 m
Temperature
Operating ........................................ −10 °C to +50 °C
Storage ........................................... −40 °C to +60 °C
Temperature coefficient .................... 0.1 X (specified accuracy / °C,
(<18 °C or >28 °C)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

179177

Table of Contents