175, 177, 179
Calibration Information

Maintenance

To prevent possible electrical shock, fire, personal injury, or damage to the Product:
Repair the Product before use if the battery leaks.
Do not operate the Product with covers removed or the case open. Hazardous
voltage exposure is possible.
Remove the input signals before you clean the Product.
Use only specified replacement parts.
Have an approved technician repair the Product.
Use only specified replacement fuses.
Replace a blown fuse with exact replacement only for continued protection
against arc flash.

Clean the Product

To avoid electrical shock:
Remove test leads and any input signals before cleaning.
Do not reinstall the pca until it is completely dry.
Dirt or moisture in the terminals can affect readings.
To avoid damaging the Product:
Do not use aromatic hydrocarbons or chlorinated solvents for cleaning. These
solutions will react with the plastics used in the instruments.
Do not use detergents of any kind for cleaning the pca.
Clean the instrument case with a damp cloth and mild detergent.
The pca may be washed with isopropyl alcohol or hot deionized water and a soft brush. Remove excess
cleaning material with clean dry air at low pressure, then dry the pca at 50 C.
8
Figure 2. Battery and Fuse Replacement
 Warning
 Warning
 Caution
aik13f.eps