Duty Cycle Performance Test; Duty Cycle Test - Fluke 867B Service Manual

Hide thumbs Also See for 867B:
Table of Contents

Advertisement

867B,863
Service Manual

Duty Cycle Test

Rotate the GMM selector to AC Volts ( ). Check that 'Full Auto' is selected in the
display above x. Select Hz and Duty Cycle. Connect the GMM and the PM5139 in
Configuration 10 (Figure 4-6) and perform the test shown in Table 4-12.
Range
AUTO
1
DISPLAY
MODE
WAKE UP
4-14
Table 4-12. Duty Cycle Performance Test
Input
50.00% 10 kHz, 0.2V p-p
square
2
3
4
5
FREEZE
HZ
SAVE
MIN
RANGE
PRINT
MAX
V
TOUCH
HOLD
V
mV
HiZ
A
mA
A
A
mA/ A
EXT TRIG
COM
V
Figure 4-6. Configuration 10 (PM5139)
Input Source
Lower Limit
PM 5139
PM 5139 function generator
0.1 mHz
10MHz
LOCAL
ADDR
RESET
Upper Limit
49.8
50.2
RANGE
x10
OUTPUT
Zo 50
4-12.
os33f.eps

Hide quick links: