Resistance And Conductance Specifications; Capacitance And Diode Function Specifications; Frequency, Duty Cycle, & Pulse Width Specifications - Fluke 87 IV Series Service Manual

True rms multimeter
Hide thumbs Also See for 87 IV Series:
Table of Contents

Advertisement

87 & 89 Series IV
Service Manual

Resistance and Conductance Specifications

Function
Resistance
Conductance
1.
For the 5,000 count mode, divide the number of least significant digits (counts) by 10.
2.
For relative humidity greater than 70 %, resistance accuracy is 0.5 % over 1 MΩ and 2.5 % over 10 MΩ.
3.
In RF field of 3 V/m, add 45 counts from 470 to 1000 MHz.
4.
Using Relative Mode (REL) to zero residual reading.

Capacitance and Diode Function Specifications

Function
Capacitance
Diode Test
For film capacitor or better, using Relative mode (REL ∆) to zero residual on 1.1 nF range.
1.
2.
For the 5,000-count mode, divide the number of least significant digits (counts) by 10.
Frequency, Duty Cycle, & Pulse Width Specifications
Function
Frequency
Duty Cycle
Pulse Width
1.
Reading will be 0.00 for signals below 0.5 Hz.
2.
Duty cycle and pulse width operate on repetitive waveforms at 14.5 Hz or greater.
8
Range
1
500.00
5.0000 k
50.000 k
500.00 k
5.0000 M
30.000 M
50.00 nS
Ranges
1
1.100 nF
11.00 nF
110.0 nF
1.100 µF
11.00 µF
110.0 µF
1,100 µF
11.0 mF
50.00 mF
2
3.1000 V
Range
500.00 Hz
5.0000 kHz
50.000 kHz
999.99 kHz
2
10 to 90 %
2
5.0000 ms
50.000 ms
Resolution
0.01
0.0001 k
0.001 k
0.01 k
0.0001 M
0.001 M
0.01 nS
Resolution
0.001 nF
0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF
1 µF
0.01 mF
0.01 mF
0.0001 V
Resolution
± (0.005 % + 1)
1
0.01 Hz
0.0001 kHz
0.001 kHz
0.01 kHz
± (0.12 x voltage range / input voltage x 100 %)
0.1 %
± (0.0001 ms + 1)
0.0001 ms
0.001 ms
Accuracy
3, 4
0.05 % + 10
0.05 % + 2
0.05 % + 2
0.05 % + 2
2
0.15 % + 4
2
1 % + 4
1 % + 10
Accuracy
1 % + 5
2 % + 2
Accuracy

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

89 iv series

Table of Contents