Performance Tests

The following procedures ensure that the 860 Series GMM meets or exceeds all
customer specifications. To successfully perform these tests, it is important that you have
read the User Manual and know how to make the measurements each specific test calls
for.

mV DC Test

Connect the GMM and 5700A in Configuration 1 (Figure 4-1). Then perform the three
steps called for in Table 4-2. Before proceeding to the next test, place the 5700A in
Standby.
Step
1
300.00 mV
2
300.00 mV
3
3000.0 mV
4-4
Table 4-1. Recommended Test Equipment
Equipment Description
Fluke 5700A Calibrator
The 5700A cannot be used to calibrate the 300Ω range,
which must be calibrated at near full scale.
Fluke 5725A Amplifier
PM5139 Function Generator
Decade Capacitor, GenRad 1412-BC
Capacitance: 0-1000 µF, Accuracy ±0.25%
Decade Resistance Source, General Resistance Inc.
Model RDS 66A (±0.0125%)
Fluke 85 DMM
Fluke 5500A Multi Product Calibrator
Table 4-2. mV DC Performance Test
GMM
Input
Range
Short
15.00 mV
-2900.0 mV
Alternative Test Equipment
Model 867B
Input Source
Lower
Limit
Shorting Bar
-0.02
5700A
14.98
5700A
-2900.9
Usage
ACV, DCV, ACA, DCA, Ohms
ACA, DCA
Frequency, Duty Cycle, Logic,
Glitch Capture, External
Trigger
Capacitance,
10 µF range
300Ω range
Component Test
ACV, DCV, ACA, DCA,
Ohms, Capacitance
Model 863
Upper
Lower
Limit
Limit
0.02
-0.02
15.02
14.97
-2899.1
-2901.4
4-4.
4-5.
Upper
Limit
0.02
15.03
-2898.6