Analog Out Buzz; Completion - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Analog Out Buzz

Completion

4-22
Configuration: Boot Code settings
1. Activate Boot Code
SET CONFIGURATION menu option.
2. Select the IntelliRate menu option.
3. Select enabled to activate this option, or disabled to deactivate this
option.
4. Select Exit and wait for the patient monitor to display the message,
"Writing settings to Flash config mem..."
5. Manually reboot the patient monitor to implement this change.
6. Refer to the operator's manual to configure the patient monitor to
use this feature.
Select Analog Out Buzz to identify an ECG or an invasive BP "leads off"
condition via a supported analog out device. The default is 0 volts.
1. Activate Boot Code
SET CONFIGURATION menu option.
2. Select the Analog Out Buzz menu option.
3. Select 60Hz sine wave to display a 60 Hz sine wave, or select 0 volts
to display a flat line.
4. Select Exit and wait for the patient monitor to display the message,
"Writing settings to Flash config mem..."
5. Manually reboot the patient monitor to implement this change.
The patient monitor is now ready for normal operation. At this time,
perform procedures in
"Functional and electrical safety checks"
page 8-1.
Dash 3000/4000/5000
("Boot Code settings" on page
("Boot Code settings" on page
4-18) and select
4-18) and select
on
2000966-542D