Analog output and defibrillator synchronization tests

Equipment
Setup
8-40
Functional and electrical safety checks: Functional Checkout procedures
Use the following equipment for these tests:
Oscilloscope, or equivalent
5.5 digit precision voltmeter (HP34401A or equivalent). Needed for
calibration only.
Analog output cable (2000633-001 or equivalent)
1. Use the figure below as a reference for connecting test devices to the
DEFIB SYNC connector, located on the back panel of the patient
monitor, for performing these tests.
PIN number
1
2
3
4
5
6
7
8
2. Test the ECG, Arterial BP, and Marker Out signals from the DEFIB
SYNC connector. ECG Input voltage of 1 mV = 1 V analog output +-
Dash 3000/4000/5000
Analog output cable pinout (2000633-001)
Black
Green
Red
Brown
Blue
White
Yellow
Gray
039B
Color
2000966-542D