Preventive maintenance: Cleaning, disinfecting and storing GE ECG cables and leadwires

Sterilization

Cautions

Storage

Improper cleaning products and processes impact or results

5-8
NOTE
EtO sterilization is not recommended, but may be required for cables
and leadwires. Frequent sterilization will reduce the useful life of
cables and leadwires.
Sterilize with ethylene oxide gas (EtO) at a maximum temperature of 50°
C/122° F. After EtO sterilization, follow the recommendations from the
sterilizer manufacturer for required aeration.
Never immerse the handheld device, cables, or leadwires in any
liquid.
Do not pour or spray any liquid directly on cables or leadwires or
permit fluid to seep into connections or openings.
Never use conductive solutions, solutions that contain chlorides, wax,
or wax compounds to clean handheld devices, cables or leadwires.
Never use solutions or products that contain the following:
Any type of Ammonium Chloride such as, but not limited to:
Dimethyl Benzyl Ammonium Chloride or Quaternary
Ammonium Chloride solutions
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Betadine
Alcohol-based cleaning agents
Sodium salts
Never autoclave or steam clean cables or leadwires.
Store in a dry well-ventilated area.
Vertically hang cables and leadwires.
Do not coil leadwires or cables tightly around any medical device.
Product discoloration
Metal part corrosion
Brittle wires
Brittle and breaking connectors
Reduced cables and leadwires life
Unit malfunction
Void warranty