Useful Error Data - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Useful error data

Process Name
Error Code
6-14
Troubleshooting: Review errors
Below is sample error log followed by a description of parameters found
in the error log.
The name of the software task that was operating when the event/
problem occurred.
The error log contains more than just operating system errors. Many
events that have an impact upon the system are also entered into the log.
The 700-series of error codes are system-initiated events. Listed below
are some of the event/error codes you might find useful.
Error code
400-4FF
Network errors were detected.
703
Diagnostic tests were completed.
70B
Internet address was changed. The network address for the patient
monitor was changed. This should only be done by qualified service
personnel.
70C
Video test was completed. This test should only be performed by
qualified service personnel.
70E
Time was changed from this patient monitor. This helps determine
how the system-wide time may have been altered.
Dash 3000/4000/5000
This information is only
useful to the manufacturer.
Definition of error codes