Required Characteristics And Configuration Of Network For Support Of Dash Patient Monitor - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Ping - MC Network
Communication partner
Device/IP address/network
Middle layer protocols
Application layer protocol and encoding
Ports
Traffic characterization and bandwidth requirements
Software transfer
Network connection on device
Usage type/function/purpose
Licensed/optional/required
Communication partner
Device/IP address/network
Middle layer protocols
Application layer protocol and encoding
Ports
Traffic characterization and bandwidth requirements
Required characteristics and configuration of network for support
of Dash patient monitor
2000966-542D
Network disclosure to facilitate network risk management:
The network must meet the specific requirements above for all traffic
flows associated with the subset of features, use cases and workflows
required by the responsible organization's users.
In addition, the network must be "flat" (i.e. limited to a single IP
broadcast domain).
The maximum MC Network size (number of unique RWhats
generated) supported by Dash is 1023.
Dash 3000/4000/5000
PC, Other Medical Devices/ MC Network
ICMP
N/A
N/A
On-demand. No ping request can be run from device. Multiple
requests could be received from other devices. Size is determined
by sending device.
MC Network
Device Servicing / Perform Software Upgrades
Required
PC, Central Station/ MC Network
UDP
Unity
Standard Unity ports
On-demand. User-initiated. Size varies by software to be
transferred.
Value
Value
C-5