Upper OP Panel/Touch Panel Assembly installation
Table 8-96
1.
Place the Upper OP Panel/Touch Panel
Assembly in the frame.
Make sure the position the cable grounds
shielding is under the Grounding Strap
Clamp when it is secured. Also, the USB
cables have markings and should be in the
same area. Notice that the cables are not
crossed or snug when routed into the
Lower OP Frame Assembly.
Apply the Grounding Strap Clamp loosely,
then connect cables. Adjust lengths as
necessary and tighten the Clamp as
shown.
2.
Connect the cables to the Operator Panel, Upper. See:
Panel assembly cables - R4 and earlier" on page
Do not stretch on the Ribbon Cable. If stretched, the connector on Operator Panel, Lower may
break, resulting in a malfunction.