Fig.4
Fig.5
Fig.6
3.The pump stop inflating and pressure begins to decrease gradually,
during which the user's blood pressure and pulse will be calculated as
Fig.6;
4.There will be a long beep following the accomplishment of measure-
ment. The air in the cuff will deflate quickly and the blood pressure
reading, pulse reading will show in the display. Moreover the measur-
ing time will also display together in two screens alternately. At the
same time, the 'M1' or 'M2' will flash to remind the user to record the
reading as Fig.7;
Fig.7
Fig.8
5.Press button 'M1' or button 'M2' to record the reading in correspond-
ing memory. For example, if button 'M2' is pressed, the display will
show as Fig.8. If the user does not press button 'M1' or 'M2' or press
button 'O/I', the reading won't be recorded;
6.Press the button 'O/I' to return to the standby mode. Please rest for
at least 3 minutes for another measurement. The device keeps unused
for 3 minutes, the device will be returned to standby mode automati-
cally.
AUTOMATIC INFLATION
There are 4 given levels of given inflation pressure for this device:
190mmHg, 230mmHg, 270mmHg and 300mmHg. When 190mmHg is
not enough or movement of arm occurs, the device will automatically
inflate to reasonable pressure level to ensure a successful measure-
ment. It is not a fault.
11