Symbols | Meaning |
Manufacturer | |
Authorized Representative in the | |
Symbol for the marking of electrical and electronics devices according to Directive 2002/96/EC. The device, accessories and the packaging have to be disposed of waste correctly at the end of the usage. Please follow Local Ordinances or Regulations for disposal. | |
CE marking in conformity with EC | |
Keep dry | |
Attention, consult accompanying manual | |
Type B Applied Part |
This instruction manual is intended to assist the user for safe and efficient operation of the wrist type digital blood pressure monitor (hereinafter: device) model LD-733. The device must be used in accordance with the procedures described in the manual. It is important to read and understand the entire manual, especially the section <Tips on taking blood pressure measurement>.
This device is intended for the non-invasive measurement of systolic and diastolic arterial blood pressure and pulse rate in adults (age 15 and above). Consult the physician if measurement is taken in children or persons with arrhythmia as errors may occur.
This device adopts the oscillometric technology with Fuzzy Algorithm measuring the arterial blood pressure and pulse rate. The wrist cuff is wrapped around the wrist and automatically inflated by the air pump. The sensor of the device catches weak fluctuation of the pressure in the wrist cuff produced by extension and contraction of the artery of the wrist in response to each heartbeat. The amplitude of the pressure waves is measured, converted in millimeters of the mercury column, and is displayed by digital value. Annotation: This device can not provide reasonable accuracy if used or stored in the temperature or humidity beyond the range stated in the section <SPECIFICATIONS> of this manual.
Fuzzy Algorithm is the processing algorithm taking into account of the specialty of individual heartbeats, which provides higher accuracy of measurement.
Sit at the table and let the table support your arm as you take the measurement. Make sure that the wrist is at approximately the same level as the heart, and that the forearm is extended naturally with palm up on the table. Wrap the cuff around the left wrist as shown in the illustration, and make sure that the wrist cuff is attached to the wrist securely with display facing the user. If your sleeve restricts blood circulation of your wrist or you roll your sleeve up so as to result in such restriction. Please take off your clothes to get accurate measurement if necessary. If measurement on your left wrist is difficult, you can use right wrist for measurement.
There are 4 given levels of inflation pressure for this device: 190mmHg, 230mmHg, 270mmHg and 300mmHg. When 190mmHg is not enough or movement of wrist occurs, the device will automatically inflate to reasonable pressure level to ensure a successful measurement. It is not a fault.
If you do not feel well during measurement or want to stop the measurement for some reason, you can press the O/I button. The device will quickly release the air in wrist cuff and the device will be returned to standby mode.
After a measurement is finished or when the device stands by, hold down button 'M' for at least 5 seconds, the display will show 'CLR' which means the stored reading is removed as Fig.7.
Standards for assessment of high or low blood pressure, regardless of age, have been established by World Health Organization (WHO) as shown in the chart as below: For example, if your blood pressure is 140mmHg (Systolic Pressure), 76mmHg (Diastolic Pressure), according to the world health organization standard, your blood pressure level is Mild Hypertension
Arterial blood pressure classification by World Health Organization
Note:
If the systolic blood pressure and diastolic blood pressure fall into different categories, the higher value should be taken for classification.
INDICATION | POSSIBLE REASON | CORRECTION METHODS |
The wrist cuff is put on wrongly. Movement of arm (or hand) or talking during measurement. The wrist cuff is not inflated to necessary pressure. | Make sure that wrist cuff is put on correctly and repeat the measurement. Repeat the measurement with following completely recommendations of manual. Repeat the measurement with pumping wrist cuff to higher pressure | |
The batteries are weak | Replace all 2 batteries with new ones |
Do not modify the equipment without authorization of the manufacturer.
SYMPTOM | CHECK POINT | REMEDY |
No display when connect to power on | The batteries have run down The polarity of battery is wrong The contact of battery compartment is polluted | Replace all the batteries with new ones. Install the battery correctly. Clean the battery terminals with dry cloth. |
Inflation stops and re-inflate later | The automatic inflation for ensuring correct measurement Did you talk or move your arm (or hand) during measurement? | See<AUTOMATIC INFLATION> Keep quiet and silent during the measurement |
The reading is extremely low or high. | Is the wrist cuff at the same level as the heart Is the wrist cuff wrapped right? Did you strain your arm (or hand) during measurement? Did you talk or move your arm (or hand) during measurement? | Make sure that your posture is right Wrap the wrist cuff correctly Relax during measurement Keep quiet and silent during the measurement |
Pulse rate is too low or too high | Did you talk or move your arm (or hand) during measurement? Did you make measurement right after exercise? | Keep quiet and silent during the measurement Take measurement again after resting for more than 5 minutes |
The batteries are run down soon | Faulty batteries are used | Suggest to use alkaline batteries of known manufacturers |
The device is automatically turned off | It is the result of automatically turn off system | This is to save the power consumption of the device, and it is not a fault. |
Model | LD-733 |
Size | 62(L) ×61(W) ×25(H)mm |
Weight | Approximately 110g without batteries |
Measuring method | Oscillometry |
Measuring range | 40 to 260 mmHg (blood pressure) 40 to 160 beats/minute (pulse rate) |
Measuring accuracy | ± 3 mmHg for static pressure ± 5% of the reading for the pulse rate |
Inflation | Automatic by the pump |
Rapid deflation | Automatic electronic valve |
Batteries | 2"AAA"×1.5V |
Memory | 90 sets of memory |
Operation temperature and humidity | +10℃ to + 40℃, 85% and below |
Storage temperature and humidity | -20℃ to + 50℃, 85% and below |
Cuff size | Applicable for wrist size 12.5-20.5cm |
Complete kit | Main body, wrist cuff, 2×AAA batteries(optional), instruction manual, warranty card |
SHANGHAI INTERNATIONAL TRADING CORP. GMBH (HAMBURG)
Address: Eiffestrasse 80, 20537 Hamburg, GERMANY
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Scian LD-733 - Wrist Digital Blood Pressure Monitor Manual