SECTION 4
GPIB Interface
Table 4-8.
Model
393OA Sweep Parameter
Setting
Messages
ERJZQUENCY
FREQUENCY
FREQUENCY
FREQUENCY
O.lE-03 (O.lmHz) to
Resolution:
O.lmHz
SNF l.OE+2 (100Hz)
SWEEPWJQ@R
FREQUENCY
SWEEP START PERIOD
SWEEP STOP PERIOQ
MKF
NR3
STD
NR3
SPD
NR3
Sets sweep marker frequency (frequency:
Yes
Hz).
Range:
O.lE-03 (0.lmI-W
to
1.2E+O6 (1 .?.MHz)
Resolution:
O.lmHz
Example:
MKF l.OE+2 (100Hz)
Sets sweep start frequency by period
Yes
(p+xizs).
Range:
0.86E-06 (0.86 ps = 1.19MHz)
to lO.OE+O3 (lOC4YJs = O.lmHz)
Resolution:
0.01 ps
Example:
STD l.OOE+OO (1s = 1Hz)
Sets sweep stop frequency by period
Y'S
(period: s).
Range:
0.86E46 (0.86 ps = 1.19MJh)
to lO.OE+03 (l@OOOs = O.lmHz)
Resolution:
0.01~s
Example:
SPD l.OOE+OO (1s = 1Hz)