Operating Examples - Keithley 3930A Operator's Manual

Hide thumbs Also See for 3930A:
Table of Contents

Advertisement

SECTION2
Getting Started
'igure
2-2.
Model 3930A
Typical Connections
FCTN
out
2.5.3
Operating
Examples
The following
examples give step-by-step
instructions
for setting basic Model 3930A operating parameters. The
Model 3930A can be connected to an oscilloscope to view
the results of the various operating examples.
Example 1: Selecting the Waveform
Type (Function)
The waveform
type can be selected using the FCTN key
as follows:
1. Press FCTN and note that the instrument
displays
the current function
and the available functions.
2. Press the number key corresponding
to the desired
function
(O-51, or rotate the MODIFY
knob until the
desired function number is displayed.
For example,
press 3 to select the square-wave
function.
The
waveform
will immediately
change to the selected
function.
3. Press DSPL to return to the normal display mode.
Example 2: Setting fhe Waveform
Frequency
or Period
Use FREQ to set the frequency or period of the output
wavefoml
as follows:
1. Press FREQ to enter the frequency-programming
mode. The instrument
will display the allowable fre-
quency range (O.lmHz to 1.2MHz for sine waves,
O.lmHz to 1OOkHz for all other waveform
types).
2. To directly
enter a completely
new numeric
fre-
quency value, key in the desired number followed
by the appropriate
units key (Hz or kHz). For exam-
ple, toenter afrequencyof
lO.lkHz,press:
1 0. 1kHz.
3. To simply modify an existing frequency value, place
the cursor on the digit to be changed using 4 or .,
then use the MODIFY knob to set the digit to the de-
sired value. Repeat as necessary for all digits to be
changed. Note that you can press the STEP SIZE key
to change the increment, or to multiply
or divide by 2
or by 10.
4. To display
the time period
of the waveform
fre-
quency, press the set key. You can then key in a new
time period or modify the existing period, if desired.
5. Press Hz or kHz to return to frequency display.
6. Press DSPL to return to normal display.
Example 3: Setting the Output Amplitude
Use the AMFTD key to set the output voltage amplitude
as follows:
1. Press AMPTD, and note that the instrument
displays
the current amplitude
and allowed amplitude
range.
2. To enter a completely
new amplitude
value in p-p
units, key in the numeric value, then press V or mV,
as required. For example, to enter a 30mV p-p ampli-
tude, press: 3 0 mV.
3.
To simply modify the existing amplitude
value, use
the cursor keys and the MODIFY
knob to set the
value as required.
4. Press DSI'L to rehun to normal display.