8. What do I do if...
... no weight display comes on?
– Is the scale switched on?
– Check the mains connection and the connection of the battery pack
– Is the plug contact on the scale correctly connected?
– Is the plug contact for connecting the display to the scale correctly connected?
0. 0
...
does not appear before the weighing operation?
– Press the Start key again – there must not be any load on the scale – and only its feet
should be in contact with the floor.
... one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.
... the display shows
– Battery voltage is dropping. Charge the rechargeable batteries in the next few days.
batt
...
appears in the display?
– The rechargeable battery is so severely discharged that weighing is no longer possible.
Recharge the battery.
St0p
...
appears in the display?
– Maximum load has been exceeded.
... the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
0. 0
and wait until
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the sca-
le to adapt to the ambient temperature and then weigh again.
E
... the display
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, interrupt the power supply by briefly removing the rechargeable
battery. If this measure is equally unsuccessful, inform the Service department.
22
?
is displayed, then weigh again.
appears?