Seca 676 Manual
Hide thumbs Also See for 676:
Table of Contents

Advertisement

TABLE OF CONTENTS

English
1. Signed and sealed . . . . . . . . . . . 52
2. Device description . . . . . . . . . . . 53
2.1 Congratulations! . . . . . . . . . . . 53
2.2 Intended use . . . . . . . . . . . . . 53
2.4 User qualification . . . . . . . . . . 54
Assembly . . . . . . . . . . . . . . . 54
Operation . . . . . . . . . . . . . . . 54
3. Safety information . . . . . . . . . . . 55
instructions for use . . . . . . . . . 55
Avoiding infections . . . . . . . . 57
Avoiding injuries . . . . . . . . . . . 57
4. Overview . . . . . . . . . . . . . . . . . . . 61
4.1 Controls . . . . . . . . . . . . . . . . . 61
on the rating plate . . . . . . . . . 63
4.5 Menu structure . . . . . . . . . . . . 65
5.1 Scope of delivery . . . . . . . . . . 66
Setting up the railing . . . . . . . 67
housing . . . . . . . . . . . . . . . . . 68
Final steps . . . . . . . . . . . . . . . 68
50 •
6. Operation . . . . . . . . . . . . . . . . . . 70
Moving the scale . . . . . . . . . . 70
Aligning the scale . . . . . . . . . 70
6.2 Weighing . . . . . . . . . . . . . . . . 71
procedure . . . . . . . . . . . . . . . 71
(TARE) . . . . . . . . . . . . . . . . . 72
mass index (BMI) . . . . . . . . . 73
to wireless receivers . . . . . . . 74
BMI automatically . . . . . . . . . 75
automatically (AClr) . . . . . . . . 77
additional weight (Pt) . . . . . . . 77
(Ahold) function . . . . . . . . . . . 78
Setting filtering (Fil) . . . . . . . . 79
(RESET) . . . . . . . . . . . . . . . . 80

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 676 and is the answer not in the manual?

Questions and answers

Summary of Contents for Seca 676

  • Page 1: Table Of Contents

    TABLE OF CONTENTS English 1. Signed and sealed ... 52 6. Operation ....70 2.
  • Page 2 8.1 Cleaning ....88 seca wireless groups ..82 8.2 Disinfecting ....88 Channels .
  • Page 3: Signed And Sealed

    1. SIGNED AND SEALED With seca products, you are not only purchasing tech- nology which has been perfected over 100 years, but also quality with official legal certification confirmed by institutions. seca products comply with European directives, standards and national laws. When you buy seca, you buy the future.
  • Page 4: Device Description

    676 purchased a highly precise and simultaneously robust device. For over 170 years, seca has used its experience in the service of healthcare and, as a market leader, it has always set standards in many countries of the world with innovative developments in weighing and measur- ing.
  • Page 5: User Qualification

    Via the wireless network, measur- seca 360° wireless ing results can be transmitted wirelessly to a seca wire- less printer or to a PC equipped with seca analytics PC software and the seca USB wireless adapter. can be moved on casters.
  • Page 6: Safety Information

    English 3. SAFETY INFORMATION 3.1 Safety information in these instructions for use DANGER! Used to identify an extremely hazardous situa- tion. If you fail to take note of this information, serious irreversible or fatal injuries will occur. WARNING! Used to identify an extremely hazardous situa- tion.
  • Page 7 Only have serv- icing and repairs performed by an authorized seca service partner. You can find service partners in your area at www.seca.com or by sending an e-mail to service@seca.com. • Only use original seca accessories and spare parts, otherwise seca will not grant any war- ranty.
  • Page 8: Preventing Electric Shock

    English • The actual transmission output of HF equip- ment may require minimum distances of more than 1 meter. Details can be found at www.seca.com. Preventing electric WARNING! shock Electric shock • Set up the device so that the power supply socket is easy to reach and the device can be disconnected from the power supply quickly.
  • Page 9: Preventing Device Damage

    • Make sure that the patient steps onto the weighing platform slowly and securely. WARNING! Danger of slipping • Ensure that the weighing platform is dry before the patient steps onto it. • Ensure that the patients feet are dry before he or she steps onto the weighing platform.
  • Page 10: Handling Measuring Results

    • Before you save measurements aquired with this device and process them for further use (e.g. in the PC software or in a seca analytics hospital information system), make sure the measurement values are plausible. • If measurement values are transmitted to the...
  • Page 11: Handling Packaging Material

    Handling packaging WARNING! material Danger of suffocation Packaging material made of plastic foil (bags) is a choking hazard. – Keep packaging material out of reach of children. – In the event that the original packing material may not be available anymore, only use plastic bags with security holes in order to reduce the risk of suffocation.
  • Page 12: Overview

    English 4. OVERVIEW 4.1 Controls Control Function Central control and display element, can be swiveled Display housing 180° to left and right Switch scale on and off Arrow key • During weighing: - press briefly: activate hold function - press and hold: activate tare function tare •...
  • Page 13: Symbols In The Display

    Control Function Arrow key • During weighing: - press briefly: activate BMI function - press and hold: call up menu • In the menu: - select submenu, select menu item - reduce value Enter key During weighing (if wireless network is set up): - press briefly: send measuring result to devices ready to receive it (PC with USB wireless adapter) - press and hold: print out measuring result (wireless...
  • Page 14: Marking On The Device And On The Rating Plate

    • max xxx mA: maximum current consumption x-y V max. xxx mA • : note polarity of device plug use compatible seca adapter only • : operate device with direct current Do not dispose of device with household waste Overview • 63...
  • Page 15: Marking On The Packaging

    4.4 Marking on the packaging Protect from moisture Arrows indicate top of product. Transport and store in an upright position. Fragile Do not throw or drop. Permitted min. and max. temperature for transport and storage Permitted min. and max. moisture for transport and storage Packaging material can be disposed of through recycling programs.
  • Page 16: Menu Structure

    *Funknetzwerk seca 360° Wireless: wireless group • Year Raum (ID): - 1 baby scale • Month Max. drei seca Funkräume: 0, 1, 2 - 1 personal scale Time • - 1 length measuring rod Maximal-Konfiguration pro Funkraum: Reset • Hour...
  • Page 17: Before You Really Get Started

    5. BEFORE YOU REALLY GET STARTED ... 5.1 Scope of delivery USA/Japan Euro Australia Component Pcs. Scale with railing in down position Display housing with display cable Pan-head tapping screw B 3.5 x 9.5 Screwdriver Power pack with adapters (model-dependent: power pack with Euro-connector) Instructions for use, not shown 66 •...
  • Page 18: Assembling The Device

    English 5.2 Assembling the device Setting up the railing Latch Locking screws 1. Place the scale on a solid, level surface. 2. Release the latch for the railing. 3. Fold up the railing and hold it in this position. 4. Tighten up both locking screws until the railing is firm without any play.
  • Page 19: Assembling The Display Housing

    Assembling the display 1. Take the display cable out of the fixing in the railing. housing 2. Connect the display cable in the railing to the cable of the display housing. Display cable 3 x B 3.5 x 9.5 3. Push the cable carefully into the railing. 4.
  • Page 20: Establishing Power Supply

    The scale may overheat, catch fire, melt or short-circuit. – Use only original seca plug-in power packs with a 9 V or controlled 12 V output voltage. 1. If required, insert the power supply connector nec- essary for your power supply into the power pack.
  • Page 21: Operation

    6. OPERATION CAUTION! Bodily injury Perform a function check as described in the section entitled "Function check" on page 89 before using the device each time. 6.1 Making the scale operational Moving the scale 1. Disconnect the power pack connector from the power supply socket.
  • Page 22: Swiveling The Display Housing

    Switching on the scale Press the Start key. All the elements of the display are shown briefly, then SECA appears in the display. The scale is operational when 0.00 appears in the display. Starting the weighing 1. Ensure that the scale has no load.
  • Page 23: Determining Extra Weight (Tare)

    Determining extra Using the TARE function, you can prevent any extra weight (e.g. a towel or a support on the weighing weight (TARE) platform) from affecting the weighing result. ATTENTION! Incorrect measurement due to force shunts If an extra weight, e.g. a large towel, contacts the surface on which the scale is placed, weight will not be measured correctly.
  • Page 24: Determining And Rating Body Mass Index (Bmi)

    English The display flashes until a stable weight is meas- ured. The weight is then continuously displayed. symbol (non-calibratable function) and the message "HOLD" are displayed. 5. To deactivate the HOLD function, briefly press the (hold/tare) arrow key. tare symbol and the "HOLD" message are no longer displayed.
  • Page 25: Sending Measurement Results To Wireless Receivers

    If in doubt, seek the advice of a consultant. Sending measurement If the scale is integrated in a net- seca 360° wireless work, you can send the measurement results to results to wireless devices ready to receive them (e.g. wireless printer, PC receivers with USB wireless adapter) at the touch of a button.
  • Page 26: Calculating And Printing Out Bmi Automatically

    NOTE The prerequisite for this function is that the devices are registered together in one wireless group (see "The seca 360° wireless network" on page 82). 1. Perform the weighing operation. 2. Briefly press the Enter key (send/print) of the scale.
  • Page 27: Other Functions (Menu)

    6.3 Other functions (menu) Other functions are available to you in the menu for the scale. This enables you to configure the scale to suit your conditions of use perfectly. …. Reset • Autoclear • PT Weight Net Weight • Autohold •...
  • Page 28: Deleting Saved Values Automatically (Aclr)

    English 4. Confirm your selection with Enter (send/print). The current setting for the menu item or a submenu are displayed (in this case, level "0"). 5. To amend the setting or call up a different sub- menu, press one of the arrow keys until the desired tare setting (in this case, level "2") is displayed.
  • Page 29: Activating The Auto Hold (Ahold) Function

    2. Confirm your selection. The last memory used is displayed. 3. You can adopt the memory displayed or select a different memory. 4. Confirm your selection. Arrows flash in the display. The additional weight saved to the selected memory is displayed. 5.
  • Page 30: Activating Beeps (Beep)

    English 1. Select the item "Ahold" from the menu. 2. Confirm your selection. The current setting is displayed. 3. Select the desired setting: – on – off 4. Confirm your selection. You exit the menu automatically. Activating beeps You can set whether a beep is heard every time a key is pressed and when a stable weight value is achieved.
  • Page 31: Switching Unit Of Weight (Unit)

    3. Select a filtering level. – 0: no filtering – 1: moderate filtering – 2: high filtering 4. Confirm your selection. You exit the menu automatically. Switching unit of weight On non-calibrated scales, you can select the unit (Unit) (Unit) in which you want weight to be displayed.
  • Page 32 English Function Factory setting Unit of weight Wireless module (SYS) Autosend (ASend) Autoprint (APrt) NOTE The wireless module is switched off when fac- tory settings are restored. Information about existing wireless groups is retained. Wireless groups do not have to be set up again. 1.
  • Page 33: The Seca 360° Wireless Network

    The following combination of devices is possible for each wireless group: • 1 baby scale • 1 personal scale • 1 length measuring rod • 1 seca wireless printer • 1 PC with seca USB wireless adapter 82 •...
  • Page 34: Channels

    • 1: personal scale • 2: length measuring rod • 3: wireless printer • 4: PC with seca USB wireless adapter • 7: baby scale • 5, 6 and 8-12: reserved for system expansion The seca 360° wireless network • 83...
  • Page 35: Operating The Scale In A Wireless Group (Menu)

    7.2 Operating the scale in a wireless group (menu) All the functions you need to operate the device in a seca wireless group can be found in the "rF" submenu. Information for navigating in the menu can be found on Page 76.
  • Page 36 11. Switch on the device, e.g. a wireless printer, you want to integrate in the wireless group. When the wireless printer is detected, you will hear a beep. The seca 360° wireless network • 85...
  • Page 37: Activating Automatic Send (Asend)

    (APrt) printer registered to the wireless group. NOTE This function is only accessible if a seca wireless printer has been integrated in the wireless group via the "learn" function. 1. Switch on the device. 86 •...
  • Page 38: Set The Time (Time)

    NOTE This function is only accessible if a seca wireless printer has been integrated in the wireless group via the "learn" function. 1. Switch on the device. 2. Select the "Time" menu item from the "rF" sub- menu.
  • Page 39: Hygiene Treatment

    8. HYGIENE TREATMENT WARNING! Electric shock The device is not de-energized when the on/off key is pressed and the display goes out. Use of fluids on the device may cause electric shock. – Before each hygiene treatment, remove the mains cable to de-energize the device. –...
  • Page 40: Function Check

    "Troubleshooting" from page 89, you may not use the device. – Have the device repaired by seca service or an authorized specialist dealer. 10.TROUBLESHOOTING Problem...
  • Page 41 Cause / Solution • The device was unable to send measurement results to the wireless receiver (seca wireless printer or PC with seca USB wireless adapter). – Ensure that the scale is integrated in the wireless network. – Ensure that the receiver is switched on.
  • Page 42: Servicing

    (Lrn)" on page 84). 11.SERVICING On leaving the factory, your seca scale has an accu- racy of ±0.15% or better. To preserve this level of accuracy, the product must be set up with care and serviced regularly. Depending on how frequently the scale is used, we recommended servicing at intervals of 3 to 5 years.
  • Page 43: Technical Data

    12.TECHNICAL DATA Technical data seca 676 Dimensions of scale when folded up • Depth 965 mm • Width 1089 mm • Height 151 mm Dimensions of scale when set up • Depth 1150 mm • Width 920 mm • Height...
  • Page 44: Optional Accessories

    Access ramp 470-00-00-009 Wireless printers • seca 360° Wireless Printer 465 Country-specific versions • Country-specific versions seca 360° Wireless Printer Advanced 466 PC software • Application-specific license seca analytics 115 packages 456-00-00-009 seca 360° Wireless USB Adapter 456 14.SPARE PARTS...
  • Page 45: Disposal

    The warranty shall become null and void where the device is opened by persons not expressly authorised to do so by seca. We ask customers based abroad to contact their local sales agent direct in the case of warranty claims.