Important Control Solution Test Information

Use only PocketScan™ Control Solution.
Check the expiration date on the control solution vial. Do not
use if expired.
Shake the vial and discard the first drop of PocketScan Control
Solution.
Use only for three months after first opening. Record the dis-
card date (date opened plus three months) on the control
solution vial. Discard after three months.
Store the control solution tightly closed at room temperature
below 30°C (86°F). Do not refrigerate.
CAUTION:
The control solution range printed on the test strip
vial is for PocketScan Control Solution only. It is used to test
meter and test strip performance. It is not a recommended range
for your blood glucose level.
11

Advertisement