nubeca PO8201 Owner's Manual page 25

Handheld
Table of Contents

Advertisement

Problem
Segments of the SpO
pulse rate displays are
missing.
Disruption in the oximeter
performance.
Displays and indicators
are off, but a continuous
audible alarm is sounding.
Red light battery indicators
appears accompanying
the message displayed on
LED as shown below:
If you have followed the actions recommended above but the problem keeps
unsolved, please contact your local dealer for assistance.
Possible Cause
or
2
Defective LED
displays.
Electromagnetic
interference (EMI).
Problem in operation.
The batteries are low.
25
Handheld Pulse Oximeter
Possible Solution
Displayed values may not
be reliable; discontinue
use of the oximeter.
Remove the oximeter from
the EMI environment.
Turn the oximeter off and
replace the batteries. If
the problem still persists,
contact your agent for
help.
Replace the batteries.