User Programs (U1 - U21) - schwa-medico EMP 2 PRO Instruction Manual

2-channel muscle stimulato
Hide thumbs Also See for EMP 2 PRO:
Table of Contents

Advertisement

Symbols on the display during stimulation:
Rising ramp:
See illustration program no. 1
Program 21 and U 21
Stimulation type
Frequency
Rising ramp
Work phase
Therapy duration
Description
Symbols on the display:
Rising ramp:
8
User programs (U1 – U21)
Create a user program
Switch ON the EMP 2 PRO. Select a USER program with the P key (U1 – U21). Press the E key
to access the editing mode. The parameters to be changed are flashing on the display. Adjust the
flashing parameter with the ▼ and ▲ keys. Press the E key to save your changes and to access
the next parameters. Repeat the operation for all parameters (frequency, pulse width, therapy
duration and ramp/pause time). Press the E key at the end to save your changes.
Description of user program possibilities
Program U1, U3, U4, U9, U20: Dynamic stimulation
Frequency:
Pulse width:
Stimulation time:
Program U2, U5: TENS
Channel 1: gate control / Channel 2: Endorphin release
Frequency channel 1:
Frequency channel 2:
Pulse width:
Stimulation time:
Program U6: HAN
Stimulation time:
Program U7, U8 : Modulation
Max. frequency:
Min. frequency:
Pulse width:
Stimulation time:
Falling ramp:
Agonist / Antagonist
Dynamic stimulation ; channel 1 and channel 2
40 Hz
Pulse width
2 s
Falling ramp
3 s
Pause
20 min
channel 1 stimulate the agonist and channel 2 the antagonist
time delayed
Falling ramp:
Pause:
20 Hz -120 Hz
70 µs - 500 µs
1 - 99 min.
80 Hz, 90 Hz, 100 Hz, 110 Hz, 120 Hz
2 Hz, 5 Hz, 10 Hz
70 µs - 500 µs
1 - 99 min.
1 - 99 min.
10 Hz - 120 Hz
2 Hz, 3 Hz, 4 Hz, 5 Hz, 6 Hz, 7 Hz, 8 Hz
70 µs - 180 µs
1 - 99 min.
EMP2 PRO | Instruction Manual
300 µs
1 s
6 s
13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents