Program description
Both channels work with the same parameters
Rising ramp: 2 s
Work phase: 3 s
Falling ramp: 1 s
Pause: 9 s
Both channels work with the same parameters
Rising ramp: 2 s
Work phase: 3 s
Falling ramp: 1 s
Pause: 9 s
Both channels work with the same parameters
Rising ramp: 2 s
Work phase: 4 s
Falling ramp: 1 s
Pause: 8 s
Both channels work with the same parameters
Rising ramp: 2 s
Work phase: 4 s
Falling ramp: 1 s
Pause: 8 s
Both channels work with the same parameters
Rising ramp: 3 s
Work phase: 5 s
Falling ramp: 1 s
Pause: 10 s
Both channels work with the same parameters
Both channels work with the same parameters
Rising ramp: 2 s
Work phase: 4 s
Falling ramp: 1 s
Pause: 4 s
Both channels work with the same parameters
Rising ramp: 2 s
Work phase: 3 s
Falling ramp: 1 s
Pause: 6 s
Dynamic stimulation
Rising ramp:
2 s
Falling ramp.
2 s
Dynamic stimulation, channel 1 and 2 are time
delayed during the treatment
Rising ramp: 2 s ; Falling ramp: 1 s
Work phase: 3 s ;
Pause: 6 s
26
EMP2 PRO | Instruction Manual
Overview pre-set programs
Frequency / Hz
Pulse width / µs
35
150
35
300
65
250
65
300
35
250
10
180
20
180
50
180
120
150
40
300
Time / min.
20
20
20
20
20
15
15
15
20
20