Factory settings
The parameters from P1 to P21 are identical in new devices to respectively U1 to U21
Program P1 and U1
Stimulation type
Frequency channel 1 and 2 80 Hz
Rising ramp
Therapy duration
Description
is working time delayed to the other. The intensity increases within 1 s from zero to the set level
and decreases within 1 s from the set level back to zero. Since both channels work in phase,
channel 1 reaches its maximum intensity level when channel 2 is at its lowest intensity level.
Dynamic stimulation offers a comfortable massage-effect and pain relief when electrodes are
placed in the right position in the pain area. Please see the special electrodes settings for
dynamic stimulation in the chapter electrode placement.
Symbols on the display during stimulation:
Rising ramp:
Program P2 and U2
Stimulation type
Frequency channel 1
Frequency channel 2
Therapy duration
Description:
Channel 1 works in gate control mode and channel 2 in endorphin release mode.
Symbol on the display during stimulation:
Program P3 and U3
Stimulation type
Frequency
Rising ramp
Therapy duration
Description:
is working time delayed to the other. The intensity increases within 0.2 s from zero to the set level
and decreases within 0.2 s from the set level back to zero. Since both channels work in phase,
channel 1 reaches its maximum intensity level when channel 2 is at its lowest intensity level.
Dynamic stimulation offers a comfortable massage effect and pain relief when electrodes are
Low back pain
Slow dynamic stimulation
Pulse width
1 s
Falling ramp
20 min
Both channels are working with the same frequency. Each channel
Falling ramp:
Lumbosciatica
Channel 1: Gate control 80 Hz
Channel 2: Endorphin release 2 Hz
80 Hz
Pulse width
2 Hz
Pulse width
20 min
Program 2 works with 2 different frequencies on each channel.
Cervico-brachial neuralgia
Fast dynamic stimulation
80 Hz
Pulse width
0.2 s
Falling ramp
20 min
Both channels are working with the same frequency. Each channel
EMP2 PRO | Instruction Manual
150 µs
1 s
200 µs
200 µs
150 µs
0.2 s