VARIOUS SETTING (MENU/Tools)

Change of compressor delay time (Tools/Comp. Delay)

There is a starting delay time setup (Comp Delay) of A compressor /B compressor in the power failure
recovery in this product. Setting can be set from 3 to15 minutes. A compressor setup in the factory
shipment is 3 minutes. An A compressor starts first. (it is seen from the rear, the left)
A compressor starts B compressor, and starts after about one minute.
Table 13 Compressor starting delay time setup process (When the starting delay time of the
compressor is changed from 3 minutes to 5 minutes.)
Description of operation
1
2
Press the MENU button.
Press the shift key, highlight the
3
"Tools" menu window, and press the
ENTER key.
Press the shift key, highlight the
4
Date Time by pressing ENTER keys.
Select Comp Delay by pressing
5
MENU
button.
ENTER key
Figure input key is pressed in order
6
of 0, 5.
7
Press the ENTER key.
8
Press the MENU button.
The shift key is pressed, and made
to highlight Cancel of the menu
9
window, and press the
key.
Key operated
Then,
pressing
ENTER
Top screen is displayed
-----
The menu window is indicated.
MENU
"set" is highlighted.
Select Tools window is indicated.
ENTER
The Date is highlighted
MENU
Comp Delay setting window is indicated.
ENTER
0
5
Comp Delay is changed at 3 to 5.
The menu window is indicated.
ENTER
"OK" is highlighted.
Comp Delay time is memorized, and the
MENU
Select tools window is indicated.
Top screen is displayed
ENTER
31
Indication after operation
And the
And the