SD CodeFree™ BLOOD GLUCOSE MONITORING SYSTEM
4)
Compare control solution result to the
range printed on the test strip container.
If the results are not within the control range
printed on the test strip container, then
the meter and strips may not be working
properly. Repeat the control solution test.
The control solution range printed on the
test strip container is for SD Glucose Control
Solution only. It is not a recommended
NOTE
range for your blood glucose level.
m g /d L
m m o l/
L
[This is an example. Refer to the ranges
on your test strip container.]
5)
Remove the used test strip for control
solution from the meter and discard it.
[Example]
Control Range
Level M
90-140 mg/dL
5.0-7.8 mmol/L
Level H
170-240 mg/dL
9.4-13.3 mmol/L
35