How to do a control solution test
STEP 1
Insert Test Strip.
®
Insert a test strip into the microdot
meter. Be sure the black contact bars
go in first and that the "test zone" is
facing you. Push the strip in firmly. Be
sure it can go no further. The meter
turns on automatically.
The code number ( example 82) will
appear on the display.
If the code number does not match
the code number on the vial label,
code the meter following the
instructions in step 3 on page 7
STEP 2
Apply Control Solution.
Be sure the control solution is at
room temperature. Shake bottle
vigorously. Remove cap; invert bottle
and squeeze out one drop of control
solution. Place a drop on your finger
just as if it were a drop of blood. Be
sure to wash your hands before
doing this
10