When the scale is on, press
3.2 Pre-Calibration
To calibrate the scale please follow these instructions:
First remove all objects from the platform, then press
to set the display to zero. Press
"SCALE" or "LINE". "SCALE" means to calibrate the scale
sensitivity (normal calibration); "LINE" means to calibrate the
scale linearity. Press
Tare
the two kinds of calibration can be interchanged. Please
proceed to the next step below)
3.2.1 Sensitivity (regular) calibration
When calibrating the scale sensitivity, two different weight
values can be used. According to the shown calibration weight
value (ie 20.000) on the display, press
calibration weight value you want to use, then press
to confirm. The display will now show "0". In a few seconds,
the display will change back to the calibration weight value (ie
20.000). Place the correct calibration weight on the center of
the platform.
In a few seconds, the display will stop, and then show the
calibration weight value again (ie 20.000). This means
calibration is complete.
3.2.2 Linear calibration
When the display is showing "LINE", press
display "0". In a few seconds the scale will show 1/3 (or1/2) of the
weight value of it's full capacity (ie 10.000); put the corresponding
weight on the center of the platform.
weight value will be shown(ie 30.000). place
weight(30g) on the center of the platform.
will stop flashing and it will show the original calibration weight value
On/off
to turn off the scale.
Tare
the display will show
Light
to confirm (If you press
Units
, to select which
Units
Tare
Tare
,the scale will
In a few seconds, the full capacity
the corresponding
In a few seconds the display
2
(ie 30.000). Calibration is complete.
NOTE: The scale linearity has been calibrated at the factory,
this type of recalibration is not necessary unless you need to
change or alter the loadcell or adjust for certain types of
environments,
3.3 Changing Weighing Units
With the scale powered ON, , press
the following units:
g—oz—ozt—dwt—gn—ct*
,
Note that the Gempro 250 has two Carat modes. One
reads in 0.001g/0.005ct increments and the other
reads to 0.002g/0.01ct. This was done so that jewelers
can choose which accuracy they require for their use.
We suggest to use 0.002g/0.01ct when possible for
maximum accuracy. Programming instructions are
shown later in this manual.
Note: The scale has last unit memory, so when you turn it off
and back on it will show the last unit used.
3.4 Tare
With the scale powered ON, press
zero.
3.5 Backlight
Press
Light
turn on automatically.
Press
Units
on;
Press
Units
off ;
Press
Tare
states.
damage or errors.
Units
Tare
twice F1 will appear, and the light (10d) may
to select F2, this means the backlight is always
to select F3, this means the backlight is always
to confirm, the backlight can be set to different
3
to change between
to set the display to