E-5 error message
• Strong electromagnetic field (e.g.: mobile phone) or high-
intensity electrostatic discharge is interfering with the
meter. Repeat the measurement.
• Meter is damaged. Check the meter with Check-strip,
refer to "AutoSense system checking".
E-6 error message
• Batteries are running low. Replace the batteries.
• High-intensity electrostatic discharge may be affecting the
meter. Remove the batteries and keep the OK button
pressed for 15–20 seconds. Replace the batteries and
switch on the meter. If "E-6" appears again, change the
batteries.
E-7 error message
• Insufficient amount of blood was applied to the strip.
Repeat test with a bigger drop of blood.
• High-intensity electrostatic discharge may be affecting the
meter. Repeat the measurement.
• Hematocrit level of sample is too high or too low.
Repeat test with a fresh drop of blood.
• Test strip is damaged. Repeat test with another strip.