Setting Stimulation Level: Use the ↑ and ↓ controls to increase or decrease
the intensity. Do not use the PRG button. After 20 seconds the stimulation
level is electronically locked as a safety feature to prevent the patient from
unintentional increase in stimulation. To increase, just press ↓ to reduce it by
at least 1 mA, which will unlock the safety feature. Thereafter you have 20
seconds to increase to desired level, before the lock is activated again. Any
change in level or mode will unlock this safety feature.
Setting Frequency: Press PRG once and use ↑ or ↓ to increase or decrease
the frequency.
SettingPulse Width: Press PRG twice and use ↑ or ↓ to increase or decrease
the pulse width.
Setting the Treatment Time: Press PRG three times and use ↑ or ↓ to select
the desired Treatment Time. Keep pressing ↓ to set a Continuous Treatment
Time.
Setting the ON-Time: Press PRG four and use ↑ or ↓ to increase or decrease
the ON-Time.
Setting the OFF-Time: Press PRG five and use ↑ or ↓ to increase or decrease
the OFF-Time.
Setting the Ramp-Up: Press PRG six and use ↑ or ↓ to increase or decrease
the Ramp-Up.
Setting the Ramp-Down: Press PRG seven and use ↑ or ↓ to increase or
decrease the Ramp Down.
Setting Mode: Press PRG eight times and use ↑ or ↓ to toggle between
Simultaneous stimulation or Alternating stimulation between the two
channels.
Setting Waveform: Press PRG nine times and use ↑ or ↓ to toggle between
Biphasic or monophasic waveforms (AC or DC).
Compliance data: Press PRG ten times and use ↓ to display the usage time in
minutes and number of times used, when prompted by text in display. Only
stimulation levels above 5 mA are recorded as usage time. The compliance
meter can be reset by pressing ↑ to show prompt in display and thereafter
press ↓ twice to reset both parameters to zero.