Technical Specifications - Zynex Medical E-wave User Manual

Table of Contents

Advertisement

Technical Specifications

Amplitude:
Frequency:
Pulse width:
On-Time:
Off-Time:
Ramp Up:
Ramp Down:
Waveforms:
Treatment timer:
Compliance meter:
Dimensions:
Weight:
Warranty:
Page 11
0-100 mA
4-100 Hz
Factory setting is 30 Hz.
50-300 µsec.
Factory setting is 200 µsec.
0.5 to 30 seconds
Factory setting is 6 seconds.
0.1 to 60 seconds
Factory setting is 6 seconds.
0.1 to 6 seconds
Factory setting is 1 second.
0.1 to 6 seconds
Factory setting is 1 second.
Symmetrical biphasic or monophasic
selectable.
Continuous, 10-100 minutes, in 10
minutes steps.
Factory setting is continuous.
Records total usage time in minutes and
number of times used.
Can be reset.
2.5 x 5.5 x 1.0 in.
8 oz. incl. Battery.
3 Years manufacturers warranty on
materials and workmanship.
Accessories excluded.