KATflow 230 1 Safety instructions, legal requirements, warranty, return policy 1 Safety instructions, legal requirements, war- ranty, return policy 1.1 Symbols used in these operating instructions Danger This symbol represents an immediate hazardous situation which could result in a serious injury, death or damage to the equipment. Where this symbol is shown, do not use the equipment further unless you have fully understood the nature of the hazard and have taken the required precautions.
KATflow 230 1 Safety instructions, legal requirements, warranty, return policy 1.3 Warranty ● Any product purchased from KATRONIC is warranted in accordance with the relevant product documentation and as specified in the sales contract provided it has been used for the purpose for which it has been designed for and operated as outlined in these operating instructions.
The clamp-on sensors are simply attached to the out- side of pipes. The KATflow 230 uses ultrasonic signals for the measurement of the flow, employing the so-called transit-time method.
The following items are typically supplied (please refer to your delivery note for a detailed description of your scope of supply): ● KATflow 230 portable flowmeter ● Clamp-on sensors (usually one or two pairs depending on pipe sizes to be measured) ●...
KATflow 230 3 Installation 3.2 Clamp-on sensor installation The correct selection of the sensor location is crucial for achieving reliable meas- urements and a high accuracy. Measurement must take place on a pipe in which sound can propagate (see Acoustic Propagation) and in which a rotationally sym- metrical flow profile is fully developed (see Straight Pipe Lengths).
Page 10
KATflow 230 3 Installation Correct Disadvantageous For a vertical pipe: Select the measuring point at a location where the liquid flows upward to insure that the pipe is completely filled. Correct Incorrect Table 1: Recommendations for sensor mounting location Look for a sensor installation location with sufficient straight pipe to obtain accurate measurements.
KATflow 230 3 Installation Disturbance source: T-section Inlet Outlet L ≥ 50 D L ≥ 10 D Disturbance source: diffuser Inlet Outlet L ≥ 30 D L ≥ 5 D Disturbance source: reducer Inlet Outlet L ≥ 10 D L ≥ 5 D...
KATflow 230 3 Installation 3.5 Clamp-on sensor mounting configurations and separation distance Reflection Mode The most common clamp-on sensor mounting configuration is the Reflection Mode, also known as V-Mode (see Illustration 3, sketch (1). Here the ultrasonic signal pass twice through the medium (2 signal passes). The Reflection Mode is the most convenient mounting method as the transducer separation distance can be meas- ured easily and the sensors can be perfectly aligned.
RS 232 serial Temperature sensor Pt100 1 system connector interface sensor Pt100 2 Drawing 3: Electrical interconnections overview KATflow 230 3.7 Clamp-on sensor mounting Before the sensors can be mounted ● the installation location should have been determined, ● a sensor mounting method should be chosen, ●...
KATflow 230 3 Installation Diagonal mode Reflection mode A<0 A= Transducer distance 080114_1 Illustration 4: Sensor pipe mounting configurations 3.7.2 Acoustic coupling gel In order to obtain acoustical contact between the pipe and the sensors, apply a bead of acoustic coupling gel lengthwise down the centre of the contact area of the sensors.
KATflow 230 3 Installation 3.7.3 Correct positioning of the sensors 080114_1 Illustration 6: Correct positioning of the sensors Always mount the transducer pair such that the free front edges of the sensors face each other. There is a different engraving on the top of each transducer. The transducers are mounted correctly if the engravings on the two transducers are forming an arrow to- gether.
Page 18
KATflow 230 3 Installation ● Take the spring end of the chain in one hand and insert the last ball ele- ment in the vertical slot of the retaining clip. Mount the chain around the pipe. ● Pull the chain firmly around the pipe and fasten it in the lateral slot of the retaining clip.
KATflow 230 4 Operation 4 Operation 4.1 Switching On/Off The flowmeter is switched on by the <ON> key for more than 2 seconds continu- ously. Equally it can be switched off by pressing the <OFF> key for more than 2 seconds.
KATflow 230 4 Operation 4.3 Keypad and display Illustration 9: Keypad and display overview 4.3.1 Keypad key functions Main function Secondary function Alphanumerical entry: Show NEXT available item (1 short key stroke) (2 short key strokes) (3 short key strokes)
Page 21
KATflow 230 4 Operation < Alphanumerical entry: > Alphanumerical entry: = Show positive total value Alphanumerical entry: Toggle MUltipleXer (for multi-chan- nel functionalities) Alphanumerical entry: = Stop totaliser function Alphanumerical entry: DIRECT menu access Move menu/list selection item Alphanumerical entry:...
KATflow 230 4 Operation 4.3.2 Display functions Display icons Menu header Display line 1 Display line 2 Date, time (optional) Illustration 10: Main display functions Display icon Function Icon not used Datalogger recording Datalogger switched off 1 segment = 33% battery power available...
KATflow 230 4 Operation RTC operating RTC failure Error recorded in error log No error detected Serial output RS 232 switched on Serial output RS 232 switched off 4.4 Quick setup wizard The quick setup wizard allows for a speedy setup of the most important parameters...
Page 24
KATflow 230 4 Operation Select fluid using cursor keys. Confirm by pressing <ENTER>. Enter process temperature using alphanumer- ical keys and confirm by pressing <ENTER>. Use key <UP> as character backspace clear to correct for data entry errors. Select pipe lining material using cursor keys.
KATflow 230 4 Operation 4.5 Measurements 4.5.1 Main process value (PV) display The measurements are started either through the Quick Setup Wizard or the menu structure by selecting Start Measurement upon first power on. Once all the para- meters are programmed, any subsequent power-on sequences will bring up the main PV display immediately.
KATflow 230 4 Operation 4.5.4 Totalisers The totaliser displays will only be shown when the totalisers are activated. Display screen Operation The flow totaliser can be started by press- ing <Q >. Pressing <Q > shows the total accumu- lated flow in positive flow direction.
KATflow 230 5 Commissioning 5 Commissioning 5.1 Menu structure Main menu Menu level 1 Menu level 2 Description/settings Quick Start Setup Wizard Single Units Select from list ↑↓ in/s m3/h m3/min m3/s l/min USgall/h USgall/min USgall/s bbl/d bbl/h bbl/min kg/h...
Page 28
KATflow 230 5 Commissioning Hydrochloric acid Sour cream Sulphuric acid Toluene Vinyl chloride User (kinematic viscosity, density, medium c- speed) Kinematic vis- Only if user fluid selected cosity 0.001 ... 30000 mm Density Only if user fluid selected 100 ... 2000 kg/m...
Page 29
KATflow 230 5 Commissioning ment Sensor type Indication of sensor type and serial number if automatically detected, otherwise select from list ↑↓ Special SP 1 ... SP 8 Only for special, unrecognised sensors Sensor place- ment Installation Pipe Material Select from pipe material list ↑↓...
Page 30
KATflow 230 5 Commissioning No – Pulse output off Pulse Value Totaliser value of selected PV at which a pulse is generated, e.g. PV = [m3/h], Pulse Value = 10, a pulse is output every 10 m3 0.01 ... 1000...
Page 31
KATflow 230 5 Commissioning User Identifier Example: Pump P3A 9 alphanumeric character string Tag No. Example: 1FT-3011 9 alphanumeric character string Test Test Mode Control system simulation: 60 second ramping up of flow velocity in m/s from 0 to programmed Max F Cut and subsequent 60 second ramping down, i.e.
KATflow 230 5 Commissioning 2400 9600 (Default) 19200 Parity Select from list ↑↓ None Even (Default) Type Select from list ↑↓ None RS232 5.2 Diagnostics Diagnostic screens can be accessed directly during measurement or through the menu structure. 5.3 Display settings Customer specific settings for data to be displayed can be achieved by accessing the appropriate menu items.
KATflow 230 5 Commissioning 5.4.1 Serial interface RS 232 The RS 232 serial interface can be used to transmit data on-line or to download the Serial interface integral datalogger content. The settings can be found in submenu Serial Comm. RS232 Analogue outputs 5.4.2 Analogue current output 0/4 ...
KATflow 230 5 Commissioning Wiring Form C (SPDT-CO) contacts Electrical V = 48 V characterist- = 250 mA 5.5 Input configurations There are two 4-wire circuit PT100 inputs available at the bottom of the enclosure. 5.5.1 PT100 inputs Inputs Wiring...
KATflow 230 7 Troubleshooting 7 Troubleshooting Should there be the need to call customer service, please let us know the following details: ● Model code ● Serial number ● SW, HW revision ● Error log list Possible error messages may include the following:...
Page 36
KATflow 230 7 Troubleshooting error call customer support COMM READ AVE Hardware Internal communication Power on/off, otherwise error call customer support COMM READ RAW Hardware Internal communication Power on/off, otherwise error call customer support COMM READ HIS- Hardware Internal communication...
Page 38
KATflow 230 8 Technical data All data given at 25 ºC (77 ºF) unless otherwise stated Change Kinematic Viscosity Sound Speed v/ºC x10-6 Chemical Specific Substance ft/s m/s/ºC m Formula Gravity Acetic anhydride (CH3CO)2O 1.082 (20 ºC) 1,180 3,871.4 0.769 8.274...
General of liquids and liquified gases in full pipes. The KATflow 230 is designed for portable use to meet the Measuring principle : Ultrasonic time difference needs of the Service/Maintenance and Commissioning correlation principle Engineer wishing to check the flow rate at different Flow velocity range : 0.01 ...
Page 44
KATflow 230 9 Specification Flowmeter (cont.) Digital (Open- Collector) : Totaliser, value 0.01 ... Power 1000/unit, width 30 ... 999 ms, consumption : < 5 W U = 24 V, I = 4 mA Signal damping : 0 ... 99 s...