Summary of Contents for Gardsman IC2020

  • Page 1 02-April-2010 For INSTANT CARE...
  • Page 2: Table Of Contents

    4.1.1. Auto-Answer Incoming Phone Calls__________________________________________________ 22 4.1.2. Nurse Call _________________________________________________________________________ 22 4.1.3. AC Power Check Up ________________________________________________________________ 22 4.1.4. IC2020 Low Battery _________________________________________________________________ 22 4.1.5. Devices Low Battery ________________________________________________________________ 23 4.1.6. Automatic Check-in Report = Periodic Test Call_______________________________________ 23 4.1.7. Mobility Timer ______________________________________________________________________ 23 4.1.8.
  • Page 3 5.3. Scancom Event code___________________________________________________________ 29 5.4. IC2020 Programming Command Table ___________________________________________ 31...
  • Page 4: Application Overview

    WHITE NURSE BUTTON — Press once in normal mode to reset mobility timer (see command 44) — Before IC2020 dials out for alarm reporting, press once to cancel the alarm reporting Phone Jack marked — During or at-the-end of conversation, —...
  • Page 5 — DC 12 V 500 mA switching power BACK VIEW REMOTE OVERVIEW (IC2020TX) Rechargeable Battery Switch Emergency Button/LED (Red/Red) Rubber Cap — Press once to dial the Emergency line and send an Emergency alarm Mounting Bracket report to the Monitoring Station. —...
  • Page 6: The Power Supply

    7.2 V rechargeable battery. During normal operation, the AC power IC2020 adapter is used to supply power to IC2020 and at the same time recharge the battery. When the battery is fully charged, it serves as a back-up power source for a period of at least 30 hours.
  • Page 7: Line Failure Detection

    (even after powered on for one minute). 1. If line faliure lasts over 10 minuites, IC2020 will emit a beep sound and the Red LED will flash 3 times every 3 seconds. 1.5. How to install the Control 2. If line failure lasts under 10 minutes and...
  • Page 8: Learn-In The Devices

    Or, if no signal is received within 30 section 3.2 Programming Your IC2020: sec, IC2020 will exit Learning mode. Commands 91-94. IC2020 will emit a short beep and the Red LED turns off to indicate the system is back to Idle mode. Learning Mode: Learning Mode is only used to learn and 2.2.
  • Page 9 Learning Mode. Or, if no signal is received within 30 sec, IC2020 will exit Learning mode. IC2020 will emit a short beep and the LED turns off to indicate the system is back to Idle mode.
  • Page 10: System Configuration

    Local programming is prohibited when 3.1.2.2. Dial in Twice (Ring Count Disable) AC Power fails. If 00 is set in Command 41, it means IC2020 is The first digit of Access Code must be disabled to auto answer the incoming calls by entered within 15 seconds, otherwise ring count.
  • Page 11: Ctc-835 Programmer

    Step 5. Enter (#) to terminate the Command IC2020 Operation. How to enter Commands? Step 6. IC2020 will emit a short beep, indicating successful programming. Please make sure IC2020 is in Programming The maximum interval between key Mode (RED LED flashes continuously) strokes minutes.
  • Page 12 To delete the 1 Tel. number, enter (01) followed by (#), after one short beep, enter & 3 Tel. #s (23) another (#). IC2020 will emit another short beep, signaling the 1 Tel. number is & 4 Tel. #s (24) erased.
  • Page 13 #5 and/or #6. On-by-One When multiple Tel. numbers are Factory default is set as (0), Alternative. selected, the IC2020 will always dial in respective programmed order. < < >...
  • Page 14 (see Command 20-24 Command #50) up to a max. length of 40 seconds. IC2020 will emit a beep Reporting Method via the telephone handset, signaling Commands (20-24) are used to program 20 & 10 sec of time remaining to alert the Alarm reporting method for each Tel.
  • Page 15 During the Speech period, IC2020 will Audible Fault Alert emit two beeps via the telephone The IC2020 can be programmed to sound handset to alert the call recipient the detected and existent Fault Alert Beeps remaining time of 20 & 10 sec.
  • Page 16 Command 35 Function Selection Result Silent Mode IC2020 can be selected to operate in Silent (disable) Mode. This will mute all normal operation warning beeps, Guard Time tone, key tone and ring. (enable) < > < > Factory Default is set as (0), OFF.
  • Page 17 Reporting Methods (when Command #20 - automatically. #24 is set as (0), (1) or (2)) Failure to enter the correct Access Code within 2 mins will cause IC2020 Function Selection Result to automatically disconnect the call. To terminate the call, press DTMF (9) Follow-On or NURSE BUTTON on IC2020.
  • Page 18 > > (1) - (7) 1 - 7 DAYS If 00 is selected, it means the IC2020 30 DAYS cannot auto-answer the call by ring count. If you want to do Remote Factory Default is set as (5), 5 DAYS.
  • Page 19 This command is used to program the allowable time duration for Follow-On It is used to select the Follow-On Choice mode choice (Commend #45) & Two-way for IC2020 after it has made a report to the voice communication period (Speech Monitoring Center.
  • Page 20 Once enabled, Microphone. IC2020 will be able to receive the check-in Please use command #51 to check the signals from Wrist Pendants (WTR-) to recorded message. indicate their proper functioning. To re-record, please repeat Steps 1-4 to Function overwrite the message.
  • Page 21 Panic Mode Enable < > < > The IC2020 can be programmed to enable During the line failure period, the voice Panic Mode or not. prompt “Phone line trouble” will be When set as ON, Panic Alarm can be played by the set repeating cycle, but...
  • Page 22 Function Selection Result sec, IC2020 will emit a long beep automatically returns Programming mode. (01) 1 minute Step 3: IC2020 will emit 2 beeps to indicate (02) 2 minutes Pendant learnt-in successfully. (03)-(60) 3-60 minutes If one long beep is emitted instead of 2 beeps, this indicates: Factory default is set as (60), 60 minutes.
  • Page 23 Programming mode. automatically returns Programming mode. Step 3: IC2020 will emit 2 beeps to indicate the device is learnt-in successfully. Step 3: IC2020 will emit 2 beeps to indicate the Silent Pendant is learnt-in Adding a Pendant for a second successfully.
  • Page 24 Command 99 Exit Programming Mode Enter (99) and (#), after one short beep, IC2020 exits the Programming Mode and returns to idle mode.
  • Page 25: Operation

    #31 is set as ON (factory default is OFF). time, and report to the Monitoring Center. For an Incoming telephone call, IC2020 Step 4. At the last 20 & 10 sec, IC2020 will will emit cycles of 2 quick beeps. emit a beep via the telephone handset User can answer the call either by to alert the call Recipient.
  • Page 26: Devices Low Battery

    5 minutes to notify the user for 30 minutes. In addition, IC2020 will emit 4 short beeps every hour. Press the NURSE button on When the Mobility timer is not reset after IC2020 will stop the warning beeps.
  • Page 27: Remote

    Enter DTMF (3) to LISTEN-IN ONLY (about 3 secs per beep) will be heard after Enter DTMF (9) to HANG UP IC2020 dials out, to inform you that the call is in progress (regardlss of Digital or Speech Reporting Method).
  • Page 28 IC2020 will emit two beeps be played up to a max. of 40 seconds. to alert both the user and the Monitoring IC2020 will emit a beep via the telephone Center Personnel. handset signaling 20 & 10 sec time remaining to alert to the call Recipient.
  • Page 29: Walk Test (Range Test)

    This cycle repeats until all reporting is/are answered successfully. At 20 & 10 seconds before the time period expires, IC2020 will emit two beeps via the During the 10-min pause, IC2020 returns to telephone handset to alert the user.
  • Page 30: Global Test

    Step 2. Within 5 sec, press Pendant cap. button. Step 3. Use a pen or screw-driver to slide the Step 3. IC2020 will emit a short beep and Battery switch to OFF position (or the report to Monitoring Center. opposite direction of the ON marking) Step 4.
  • Page 31: Appendix

    602 - Periodic test report 147 - Sensor Supervisor Failure When the IC2020 makes periodic When IC2020 is unable to receive the Check-in reporting. signal transmitted from the Wrist 606 - Follow-On Function Transmitter for a pre-set period by Command #49.
  • Page 32: Tunstall Event Code

    D#1013 – Sensor Supervisor Failure When the Pendant 1 or Unit button is pressed. When IC2020 is unable to receive the signal transmitted from the Wrist 555155517 - Pendant 2 emergency Transmitter for a pre-set period by When the Pendant 2 is pressed.
  • Page 33 535555557 - WTR and/or PB Low battery Restore When the Low Battery status of Wrist Transmitter (WTR), Emergency Pendant (EP) or Panic Button (PB) is restored. 155555517 - SD or CO emergency When the Smoke Detector (SD) or Carbon Monoxide (CO) is pressed. 155555518 - SD or CO Low battery When the Smoke Detector (SD) or Carbon Monoxide (CO) is under Low...
  • Page 34 5.4. IC2020 Programming Command Table DTMF DTMF TERMI- DTMF CHARACTER COMMAND OPERATION DEFAULT NATOR COMMANDS Telephone number 1 1-888-852-3842 Telephone number 2 1-888-852-4098 Telephone number 3 1-866-271-5821 Telephone number 4 No number Telephone number for alarm report Telephone number for status report...
  • Page 35 DTMF DTMF TERMI- DTMF CHARACTER COMMAND OPERATION DEFAULT NATOR COMMANDS Auto answer for Remote programming, 20 rings 00 = None, 01 - 32 = Auto answer 1 - 32 rings Guard time delay 5 seconds 00 = Instant, - 99 = 99 seconds Test call, 0 = OFF, 1 = 1 day, 2 = 2 days, 3 = 3 days, 4 = 4 days, 5 = 5 days, 5 Days...