Testing with the Control Solution, Cont.
CAUTION:
The Control Solution Range is printed in the lower right of the test strip
vial label. This will allow you to di erentiate between a control solution
test and an actual blood test.
For example:
If you are using the High (Level 3) Control Solution, according to this particular
vial of strips, your meter should show a number between 255 and 345. Likewise,
if you are using the Low (Level 1) Control Solution, your meter should show a number
between 35 and 65.
Note: The control values on the test strip vial shown in the picture are not control
ranges. The control ranges of the control solutions are much greater.
ATTENTION:
1. It is recommended that glucose control solution is stored at room
temperature (46-86˚F/8-30˚C) before testing.
2. Check the expiration date before performing a glucose control
solution test.
3. Do not use if expired. Once opened, the control solution expires
after 3 months.
If your control solution test falls out of range, please follow the next
steps before contacting MHC Customer Support:
• Check the expiration dates on all the products you are using. Be sure you
are using EasyTouch® HealthPro™ Glucose Control Solution.
• Try another control solution test.
• If this test falls out of range, try another control solution test with a new
unopened bottle of strips.
• If the control test still falls out of range after following the appropriate
steps do not perform a glucose test. Please call our toll free
MHC Customer Support at (877) 358-4342.
18
Follow-up Actions
1. Check your meter and strip with the glucose control solution.
2. Perform your test again.
CAUTION:
1. If your glucose control solution falls out of range, do not perform a
glucose test and please contact MHC Customer Support
at (877) 358-4342.
2. The glucose control solution range is not the recommended range
for your blood glucose level.
3. Please discard control solutions according to your local regulations.