GUIDANCE AND MANUFACTURER'S DECLARATION
Guidance and manufacturer's declaration - electromagnetic immunity
OMRON BP710CANN is intended for use in the electromagnetic environment specified below. The
customer or the user of this OMRON BP710CANN should assure that it is used in such environment.
IEC 60601
Immunity test
test level
Conducted RF
3 V rms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Compliance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the OMRON
BP710CANN including cables, than the recommended
separation distance calculated from the equation
appropriate to the frequency of the transmitter.
Recommend separation distance
Not
Not Applicable.
Applicable.
3 V/m
d = 1.2 √P
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
32
Electromagnetic environment - guidance