GUIDANCE AND MANUFACTURER'S DECLARATION
Guidance and manufacturer's declaration - electromagnetic immunity
OMRON BP710CANN is intended for use in the electromagnetic environment specified below. The
customer or the user of this OMRON BP710CANN should assure that it is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
inputlines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
<5 % U
T
(>95 % dip in U
)
T
for 0.5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(95 % dip in U
)
T
for 5 sec.
3 A/m
Compliance level
±6 kV contact
±8 kV air
Not Applicable.
Not Applicable.
Not Applicable.
3 A/m
31
Electromagnetic
environment - guidance
Floor should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Not Applicable.
Not Applicable.
Not Applicable.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.