Twister/Twister Junior
Replace 4 hexagon bolts DIN 933 M8x20 (Pos. 2 in chapter 13.2.2) with 4 hexagon bolts
DIN 933 M8x16 and fix the lifter module to chassis again together with washers. The upper
edges of the screws must be level with the base plate.
Detach the old splash protection of the power module by loosening the two nuts.
Adhere the newly to be mounted spacers on the new splash protection for the power
module on the centre of the lower side (the side with the two holes).
Insert the cable clip to be fitted into both holes (see above).
Fix the new spray protection with two nuts. The end of the bolts must be flush with the nut
heads.
Insert the cable which leads to the lifter column into the cable clip.
Release the four adapter box screws (Pos. 6 in chapter 13.2.2).
Loosen the two inner screws in the adapter box and detach it.
Detach the cable (grey) from pin 11 and the cable (black) from pin 12 (both leading into the
lifter column) from the socket by means of a special tool.
Remove cables and box.
Attach the new adapter box to the lifter column.
Lay cables in a way so that they cannot get pinched or rub against other parts, and fix anew
with cable binders.

13.3.2 Final inspection

Activate the lift mast three times.
Check that cables cannot get pinched or rub against other parts.
Note:
The cable must be located below the external lifter pillar section. They must not be
located between the lifter pillar section and the metal plate!
Should this not be the case, the cables must be re-fixed.