Sanoclav KL Operating Instructions Manual

Sanoclav KL Operating Instructions Manual

High-pressure small-batch sterilizer, stainless steel 18/10, for sterilization in steam and water under pressure up to 2 bar = 134 °c and up to 3 bar = 143°c

Advertisement

Contact address in the event of queries:
Adolf Wolf SANOclav
Robert-Bosch-Straße 13
73337 BAD ÜBERKINGEN-HAUSEN
Tel. : (07334) 9623- 0
Fax.: (07334) 9623-12
Type of autoclave
Serial number
Year of manufacture
max. Operating pressure
max. Operating temperature :
Volume of pressure chamber :
Examined pressure
Date of examination
__________________________________________________
:
:
:
:
:
:
SMALL-BATCH STERILIZER
Type KL and TKL-MCS
STAINLESS STEEL 18/10
KL
OPERATING INSTRUCTIONS
HIGH-PRESSURE SMALL-BATCH STERILIZER
For sterilization in steam and water under pressure up
to 2 bar = 134 °C and up to 3 bar = 143°C
Edition of 07.01.2003
___________________________________________
TKL-MCS

Advertisement

Table of Contents
loading

Summary of Contents for Sanoclav KL

  • Page 1 SMALL-BATCH STERILIZER Contact address in the event of queries: Adolf Wolf SANOclav Type KL and TKL-MCS Robert-Bosch-Straße 13 73337 BAD ÜBERKINGEN-HAUSEN Tel. : (07334) 9623- 0 STAINLESS STEEL 18/10 Fax.: (07334) 9623-12 Type of autoclave Serial number Year of manufacture max.
  • Page 2 In this case the safety-valve has to be replaced unconditional pr DIN 58951 – 2 IEC 801 - 5 before the next use of the sterilizer. ADOLF WOLF SANOclav WARNING: A Cleaning or Repair of the safety-valve can only be made by the manufacturer. Bad Überkingen , 07.01. 2003...
  • Page 3 2. Preheating and after-heating temperature Preheating: If the media sensor has been activated while the „F“ key (start key) has not been de- pressed, the autoclave is preheated to an „operating readiness“ temperature of 60°C. COPYRIGHT After-heating: After sterilization the autoclave is cooled Firma Wolf Labor u.
  • Page 4 Sieve tray ∅ 17 × 3 cm – with cross perforations – double grip A 17 Acceptable risk ............2 hole, aluminium, for KL 5 and KL 7 Sieve tray ∅ 20 × 2 cm – with cross perforations – double grip A 20...
  • Page 5 7. The autoclave is heated up to the entered setpoint tem- 1.1 Explanation of symbols perature. A flashing dot at the bottom right-hand side of the display indicates when the sterilization time has started. DANGER: This symbol draws attention to a dan- CAUTION: Renewed depression of the "F"...
  • Page 6 The MCS control (microcomputer) permits digital time and tem- perature control with a PT 100 sensor and an analogue pres- WARNING: Preparation for operation: sure display. The data of the electrical heating system are either 230 V/2000 W/50 Hz (TKL-MCS 123) or 230 V/1500 W/50 Hz, Never sterilize in your autoclave : ⇒...
  • Page 7 2.6.3 Sterilization by boiling in water under WARNING: After sterilizing: pressure All steps and safety measures described in Section 2.6.1 must 1. Never open the lid while the autoclave is under be observed, together with the following additional require- pressure. The hot steam can cause serious inju- ments: ries to the hands and face.
  • Page 8 We recommend the use of our sieve tray or sieve basket (ac- cessories) to achieve uniform heating of the liquid container. WARNING: Inspection, maintenance and care: For sterilization of liquids we recommend the use of a media- 1. It is imperative that the cleaning, maintenance and sensor for autoclaves with MCS-controller (see chapter 4).
  • Page 9 B: During operation WARNING: Please observe the "Safety Instructions" IMPORTANT: The series KL and TKL autoclaves are not ad- and "Acceptable Risks" (Chapters 1. and 1.2), but mitted for sterilisation of medical products within the meaning of particularly the section "During Operation".
  • Page 10 2.6 Operation of small-batch autoclaves, 2.3 Technical data type "KL" For dimensions and weight of the individual autoclave types please refer to Table 1. Table 3: Technical data Control Parameters Heating capacity 2000 W 1500 W switching capacity closing lever a.c.