[4] Viewing Control Solution Test Result:
The result will appear on the display and, if
the volume is enabled, your meter will beep
once. The date and the abbreviation "Ctrl"
will also appear on the display.
Compare the result of your control solution
test to the range printed on the AgaMatrix
Jazz Test Strip vial label for the control
solution level (Normal or High) that you
used. Your control solution result should fall
within this range. The control solution test
result is stored in memory.
To deactivate the meter, remove
the used AgaMatrix Jazz Test Strip.
Dispose of used materials properly.
If your control solution test results
are out of the range, repeat the
test. If the results from the control
solution tests continue to fall
outside that range, do not use
the AgaMatrix Jazz System to test your blood glucose. Call
customer service.
IMPORTANT: Out-of-range results may be due to one or
more of the following factors:
[1] Expired or defective control solution.
[2] Expired or defective test strip.
20
Normal: 80-120 mg/dL
High: 220-320 mg/dL
FOR EXAMPLE ONLY.
Refer to your test
strip vial label for
corresponding range.
[3] Error in performing test.
[4] Watered-down control solution.
[5] Meter malfunction.
[6] Control solution test done outside the system operating
temperature of 50 °F to 104 °F (10 °C to 40 °C).
[7] Failure to shake the control solution bottle vigorously
before using.
[8] Failure to discard the fi rst drop of control solution and to
wipe the bottle tip clean.
CAUTION: Results from control solution tests do not refl ect
your blood glucose level. The control solution range is a
target range for control solution only. It is not a target
range for your blood glucose level.